Chapter 9
PTV Injector Maintenance
Maintaining a PTV Injector
Maintenance and Troubleshooting Manual
145
It is a good practice to change the septum with a new one every time the liner is
replaced or cleaned. The operating sequence suggested for the replacement of the
septum is reported on page 147.
When cleaning or replacing the liner
The liner must be replaced periodically, depending on the number of injections
performed and the characteristics of the samples injected. Typical symptoms will
indicate that the liner must be replaced (refer to Chapter 19,
). A dirty liner will generate tailing of sample peaks, particularly
for polar compounds, or disappearing of some peaks.
You can replace the liner with a new one or clean the liner and reinstall it.
Table 9-2 shows the available types of liners for the PTV injector.
Table 9-2.
Available liners for PTV injector
Application
P/N
Silcosteel
®
liner, 2 mm ID (provided with the GC)
453 220 44
Silcosteel
®
liner, 1 mm ID (provided with the GC)
453 220 46
Glass liner, 2 mm ID
453 220 45
Glass liner, 1 mm ID
453 220 54
Silcosteel
®
liner for simile On-column injections
453 220 52
Silcosteel
®
liner packed with quartz wool for LV injections
453 220 56
Glass sintered deactivated liner for Large Volume Injections
453 220 60
Deactivated glass liner with buffles
453 220 62
Deactivated glass liner for split and splitless injections
453 220 57
Glass liner, 2 mm ID with glass wool
453 220 70
It is good practice to replace the septum every time the liner is cleaned or
replaced. The operating sequence reported on page 151 describes how to:
•
Remove the liner and the septum
•
Clean and reinstall the liner