Installation
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 59
5.6.3
Connecting Fittings, Capillaries, and Tubing
This section provides information about how to connect and handle
capillaries, fittings, and tubing.
5.6.3.1
General Guidelines
When connecting capillaries and tubing, follow these general
recommendations:
•
Use only the capillaries and tubing (for example, solvent lines or
waste tubing) that are shipped with the product or additional or
spare capillaries and tubing as recommended by Thermo Fisher
Scientific.
•
The connectors must be free from contaminants. Even minute
particles may cause damage to the system or lead to invalid test
results.
•
Do not install capillaries or tubes that are stressed, nicked, kinked, or
otherwise damaged.
•
Install capillaries and fittings only at the positions for which they are
intended.
•
Follow the connecting instructions as required by the fitting
connection:
¨
For Viper connections, see
¨
For capillaries with conventional fitting connections (non-Viper),
see
Connecting Non-Viper Capillaries (
.
5.6.3.2
Connecting Viper Capillaries
This section describes how to connect Viper™ capillaries. The following
Viper fittings are available:
•
Viper fitting with torque toothing
•
Viper fitting with knurl
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...