7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 171
7.9
Solvent Lines and Solvent Line Filters
Follow the sequence of steps for the maintenance procedure that you
want to perform:
•
To empty the solvent lines, see
•
.
•
To replace the solvent line filter, see
Replacing the Solvent Line Filter
7.9.1
Emptying the Solvent Lines
When
To empty the solvent lines, for example, for maintenance or service
procedures or for long-term shutdown
Parts required
•
Silicone tubing
•
Syringe
Preparations
1. Turn off the pump flow and wait until the system pressure is down
to zero.
2. Prepare the pump for purging (see
Follow these steps
Follow these steps for each solvent line that you want to empty:
1. Unscrew the cap of the solvent reservoir and remove the cap
together with the solvent line from the reservoir.
2. Start a purge cycle as appropriate.
3. Stop purging as soon as the solvent line is empty.