7
•
Maintenance and Service
Page 172
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
7.9.2
Replacing Solvent Lines
To replace solvent lines, follow the sequence of steps for the solvent line
that you want to replace:
The flow path determines the pump inlet (component at which the
solvent enters the pump) and the solvent flow through the pump.
Pump Type
Pump Inlet
Binary pump
The pump inlets are the solvent selectors.
Isocratic pump
The pump inlet is the pump head.
Quaternary pump,
dual pump
The pump inlet is the vacuum degasser.
For the pump-specific operating principles that illustrate the solvent
flow through the pump, see
.
•
Solvent line from the solvent reservoir to the pump inlet: Follow the
steps
Solvent Lines From Reservoir To Pump Inlet (
.
•
Depending on the pump type
Solvent line from the pump inlet to the proportioning valve or
degasser: Follow the steps in
Solvent Lines From Pump Inlet to
Proportioning Valve or Degasser (
.
•
Depending on the pump type
Solvent line from the proportioning valve or degasser to the pump
head: Follow the steps in
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...