7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 173
7.9.2.1
Solvent Lines From Reservoir To Pump Inlet
When
Damage or blockage of the solvent line
Parts and additional items required
•
Solvent line between solvent reservoir and pump inlet
•
Solvent suitable for purging
Preparations
•
Follow these steps
1. Remove the solvent line from the reservoir cap:
a) Remove the filter from the solvent line.
b) Remove the retaining guide.
c) Pull the solvent line out of the reservoir cap.
2. Disconnect the solvent line on the pump inlet.
3. Remove the solvent line from the pump and all modules above the
pump in the system stack. When removing the solvent line from the
tubing guides, be careful not to pull on other tubing in the guides.
4. To install the new solvent line, follow the related steps in
5. To remove any air bubbles from the solvent lines, purge the pump.
7.9.2.2
Solvent Lines From Pump Inlet to Proportioning Valve or Degasser
This section applies to all pumps except the isocratic pump.
When
Damage or blockage of the solvent line
Parts and additional items required
•
Solvent line between pump inlet and proportioning valve or
degasser inlet, as applicable for the pump
•
Solvent suitable for purging
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...