7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 175
7.9.3
Replacing the Solvent Line Filter
When
•
Damage of solvent line filter
•
Impaired permeability of filter frit
Parts and additional items required
Figure 58: Solvent line filter
No.
Description
1
Solvent line filter, filter holder bottom part
2
Filter frit
3
Solvent line filter, filter holder top part
•
As applicable:
¨
Filter holder
¨
Filter frit
•
Solvent suitable for purging
Preparations
1. Turn off the pump flow and wait until the system pressure is down
to zero.
Follow these steps
1. Unscrew the cap of the solvent reservoir and remove the cap
together with the solvent line from the reservoir.
2. Remove the filter holder from the solvent line.
3.
As applicable
: Open the filter holder and remove the filter frit.