Dionex SRD-10 Installation Instructions
Doc. 065355-04 05/18
7
If the pump is a Dionex ICS-3000 DP or ICS-3000 SP, go to the next
section to update the firmware.
If the pump is not a Dionex ICS-3000 DP or ICS-3000 SP, go to
configure the pump TTLs.
Download the New Dionex ICS-3000 DP/SP Firmware
1. On the
General
tab page in the Properties dialog box, click
Install
Firmware
. A message notifies you of the firmware version that is
currently installed in the pump and the version that will be
downloaded from the data system.
If firmware version 4.1.0 (or later) is already installed, click
No
to
cancel the download. Go directly to the next section to configure the
pump TTL inputs.
2. If the installed firmware version is earlier than version 4.1.0:
a. Click
No
to cancel the download.