Dionex SRD-10 Installation Instructions
4
Doc. 065355-04 05/18
3. Place the free end of the Dionex SRD-10
WASTE
, View A)
in a waste container below the level of the workbench on which the IC system
is installed.
4. Remove the backing from the three hook-and-loop strips on the Dionex SRD-
, View B).
5. While holding the Dionex SRD-10 in the preferred orientation, press it firmly
onto the mounting surface.
6. Refer to the table below to identify the TTL input that provides on/off control
of the pump motor.
7. Locate the appropriate connector plug in the Dionex SRD-10 Ship Kit. Follow
these basic steps to connect the TTL:
a. Connect the active wire (red) and ground wire (black) from the Dionex
SRD-10 to the connector plug.
To attach a wire to the plug, strip the end of the wire (if it is not already
stripped), insert it into the plug, and use a screwdriver to tighten the
locking screw. If necessary, multiple ground wires can be attached to a
single TTL input/output ground pin.
b. Plug the Dionex SRD-10 connector plug into the connector on the rear
panel of the integrated IC system or the pump module.
Product
TTL Input
Dionex GP50, GS50, ICS-3000 SP, ICS-5000 SP*,
IP25, or IS25
TTL Input 1
Dionex ICS-1000, ICS-1500, or ICS-2000
TTL Input 3
Dionex ICS-3000 DP or ICS-5000 DP*
Pump 1: TTL Input 2
Pump 2: TTL Input 3
Dionex ICS-1100, ICS-1600, or ICS-2100
Any available TTL input
* Analytical pump only
When attaching wires to the connector plug, be careful not to allow stray
strands of wire to short to the adjoining position on the connector.