Thermo Scientific Thermo Scientific Dionex SRD-10 Installation Instructions Manual Download Page 8

Dionex SRD-10 Installation Instructions

6

Doc. 065355-04 05/18

3.1

Chromeleon 7: Pump Setup

Start the Instrument Configuration Program

1. Stop all batches, programs, and data acquisition events on the 

Chromeleon Controller that hosts the pump to be updated. 

2. If it is not already running, start the Instrument Controller Service. 

(Click 

Start > All Programs

 

> Chromeleon 7 > Services Manager

and then click 

Start Instrument Controller

.)

The Chromeleon Services Manager status message should read 
“Instrument Controller Service is running idle.”

3. Start the Instrument Configuration program. (Click 

Start > All 

Programs

 

> Chromeleon 7 > Instrument Configuration 

Manager

.)

4. If necessary, click the plus sign beside the controller name to display 

the items underneath.

5. Expand the contents of the instrument that includes the pump to be 

updated. (Click the plus sign next to the instrument.)

6. Right-click the pump to be updated and click 

Properties

.

The Properties dialog box for the pump appears. In this example, the 
pump is a Dionex ICS-3000 DP.

Summary of Contents for Thermo Scientific Dionex SRD-10

Page 1: ...Dionex SRD 10 Suppressor Regenerant Detector Installation Instructions Document No 065355 Revision 04 May 2018...

Page 2: ...ts of this document are subject to change without notice All technical information in this document is for reference purposes only System configurations and specifications in this document supersede a...

Page 3: ...tific Dionex MMS 300 MicroMembrane Suppressor that is operated in the chemical regenerant mode In the external water mode of operation an external water supply is pumped into the Dionex SRS regenerant...

Page 4: ...ot provided with the Dionex SRD 10 but is readily available elsewhere Dionex SRD 10 Ship Kit The Dionex SRD 10 Ship Kit P N 074464 includes the following parts Integrated IC Systems Pump Modules Dione...

Page 5: ...the hook and loop strips on the rear of the Dionex SRD 10 to mount it on the laboratory bench the side of the IC system or other flat surface Any orientation is acceptable Install the Dionex SRD 10 1...

Page 6: ...and ground wire black from the Dionex SRD 10 to the connector plug To attach a wire to the plug strip the end of the wire if it is not already stripped insert it into the plug and use a screwdriver t...

Page 7: ...must verify that a pump TTL input is configured in Chromeleon to control the eluent pump In addition you may need to update the pump firmware The following table indicates the setup requirements for e...

Page 8: ...ment Controller The Chromeleon Services Manager status message should read Instrument Controller Service is running idle 3 Start the Instrument Configuration program Click Start All Programs Chromeleo...

Page 9: ...SP Firmware 1 On the General tab page in the Properties dialog box click Install Firmware A message notifies you of the firmware version that is currently installed in the pump and the version that wi...

Page 10: ...rs click Yes to update the firmware 4 The download can take several seconds When the download is complete remove the firmware CD from the drive Store the CD in a safe place away from heat and moisture...

Page 11: ...Server Monitor program Click Start All Programs Chromeleon Server Monitor The Server Monitor status message should read Chromeleon Server is running idle 3 Start the Chromeleon Server Configuration p...

Page 12: ...P Firmware 1 On the General tab page in the Properties dialog box click Install Firmware A message notifies you of the firmware version that is currently installed in the pump and the version that wil...

Page 13: ...ars click Yes to update the firmware 4 The download can take several seconds When the download is complete remove the firmware CD from the drive Store the CD in a safe place away from heat and moistur...

Page 14: ...nload from the Thermo Fisher Scientific website or by contacting your local office 5 Troubleshooting Dionex SRD 10 sends TTL output signal to turn off pump although reservoir is not empty Check that t...

Reviews: