-17-
6
Connect the cooling water tube, oil circulation
tube and waste oil (drain) to the bearing housing.
Refer to the figure below for connections.
7
Put the bearing housing gently on the rotor shaft
until the bottom surface of the bearing housing
touches the door adapter. Then turn about 60
°
to
engage the stopper with the door adapter in such
manner that the water passage faces the rear of the
centrifuge.
8
Position the bearing housing stopper (B) which
prevents the bearing housing from turning in the
gap between the bearing housing and door adapter.
Water
passage
a. Connect to OIL IN
Stopper (B)
b. Connect to
DRAIN IN
c. Connect to WATER OUT
d. Connect to OIL OUT
e. Hole for water passage
Centrifuge front