-3-
1. PREFACE
We have developed a continuous flow rotor which permits sizable processing with samples
allowed to flow continuously.
Utilize this continuous flow rotor of high performance and reliability for your studies or
processing
2. SPECIFICATIONS
Table 1 shows the specifications of this continuous flow rotor.
Table 1 Specifications
Type of Rotor
TCF-32
Max. speed
32,000 RPM
Max. centrifugal force
101,685 X g
K factor
42
Total capacity
430 ml
Maximum flow rate
150 ml/min (9 l/hr)
Sample pH range
PH 4~10
Maximum Dia. of rotor
24 cm
Supply power for lubricating unit
AC 100-240V, 1A (50/60Hz)
NOTES
For operation of Model TCF-32, the following apparatus are required besides the standard
configurations of Model TCF-32 and Preparative Ultracentrifuge, which have to be prepared for
installation of Model TCF-32 at customer's site.
1. Sample Loading Pump
Specifications
Flow rate: Has to be as fast as operating flow rate required for separation of samples.
Exhaust pressure: More than 1.5kg/cm
2
.
Model without pulsating motion is recommended.
2. Low Temperature Circulating Water Cooling Bath
Cooling capacity: More than 300 Kcal/hr (water temp. at 5°C)
(Refrigerator: as strong as 300W)
Circulating Pump
Flow rate: More than 1 l/min. (pressure: 0.5-1 kg/cm
2
)
3. Density Gradient Pump
Required for density gradient separation.