Section VI Troubleshooting
Error Codes
The controller also has the capability to display error codes.
Power up Errors
Power up errors are displays until any controller key is pressed. If the error
message persists the problem is likely a keypad or controller board failure.
Display
Indication
Er 00
ROM checksum
Er 01
RAM check
Er 02
Keypad failure
Er 03
NOVRAM checksum error
Operating Errors
Operating errors are displays once and then clear after three seconds or
when any controller key is pressed. If error messages Er04 - Er14 persist,
the problem is likely controller board failure. If Er15 is displayed, ensure
serial communications is turned off in the Setup Loop. If Er16 is displayed
the unit needs calibration, contact us for assistance.
Display
Indication
Er 04 - 13
Interrupt error
Er 14
Synchronous error
Er 15
Asynchronous error
Er 16
Bad calibration
Latching Errors
In order to restart the unit when a latching error is displayed, the controller's
START/STOP button must be pressed after the condition is cleared.
Display
Indication
Lo t
Fluid temperature lower than low temp alarm setting.
Hi t
Fluid temperature higher than high temp alarm setting .
Er 22
Fluid temperature higher than the high overrange point.
This is a fixed value, factory set at 55°C.
Er 25
RDT1 input shorted.
Er 26
RTD1 input open.
LLF
Low level fault switch open for at least three seconds.
PHEr
Improper phasing.
OL
Pump motor overload.
If any other code appears contact Thermo Fisher Scientific customer service,
see Preface.