Temperature Controller

The microprocessor controller controls temperature using a PID (Propor-
tional-Integral-Derivative) algorithm. It is designed with self-diagnostic
features and easy to use operator interface.

NEXT ENTER

Use this key to accept and save changes.

YES, 

This key is used to increase numerical values.

NO, 

This key is used to decrease numerical values.

When the controller is powered it displays the reservoir fluid temperature.
Press the 

NEXT ENTER

 key to view the setpoint. The display flashes

between 

SP

 and the actual setpoint number. If desired, use the 

YES

 and 

NO

keys to change the setpoint. The display flashes as soon as either key is
depressed. Once the desired setpoint is displayed, press 

NEXT ENTER

.

When selecting an operating temperature, remember that the lowest achiev-
able temperature is a function of the available flow rate, the temperature of
the cooling water supply and the heat load.

The temperature control system actuates a control valve in the FACILITY
WATER line. The control valve adjusts the flow of the cooling water supply
to produce the desired operating temperature.

NOTE: 

The controller does not use the new value until the 

NEXT ENTER

key is depressed and the display stops flashing. The controller will not allow
you to enter a value above the maximum or below the minimum value, or
any illegal value. If you try to enter an illegal value the display will revert to
its original value when the last digit was entered.

If the 

NEXT ENTER

 key is not depressed within one minute, the controller

will time out and the new value will not be accepted. The controller will
revert to the previous value.

NOTE: 

Error codes are addressed in Section VI, Troubleshooting.

Summary of Contents for System IV

Page 1: ...ientific com tc Product Service Information Applications Notes MSDS Forms e mail Voice Info 800 258 0830 Thermo Scientific System IV Heat Exchanger Thermo Scientific Manual P N U00682 Rev 03 04 2013 I...

Page 2: ......

Page 3: ...ientific com tc Product Service Information Applications Notes MSDS Forms e mail Voice Info 800 258 0830 Thermo Scientific System IV Heat Exchanger Thermo Scientific Manual P N U00682 Rev 03 04 2013 I...

Page 4: ...uhe Germany Tel 49 0 721 4094 444 Fax 49 0 721 4094 300 info tc de thermofisher com Building 6 No 27 Xin Jinqiao Rd Shanghai 201206 Tel 86 21 68654588 Fax 86 21 64457830 info china thermofisher com St...

Page 5: ...ure 4 SECTION I Safety Warnings 5 SECTION II GeneralInformation Description 6 Specifications 6 SECTION III Installation Site 7 Facility Water Requirements 7 ElectricalRequirements 8 Voltage Selection...

Page 6: ...ion 15 Voltage Selection 15 Fuses 15 PhaseRotation 15 PumpMotor Overload Protector 16 DisplayingSoftwareVersion 16 SECTION VI Troubleshooting ErrorCodes 17 Service Assistance 18 TechnicalSupport 18 SE...

Page 7: ...hat shipped with the unit WEEE RoHS This product is required to comply with the European Union s Waste Electrical Electronic Equipment WEEE Directive 2002 96 EC It is marked with the following symbol...

Page 8: ...have a warranty against defective parts and workmanship for one full year from date of shipment See back page for more details Out of Box Failure An Out of Box Failure is defined as any product that f...

Page 9: ...leaking equipment Never operate the unit without cooling fluid in the reservoir Always turn off the unit and disconnect the line cord from the power source before performing any service or maintenance...

Page 10: ...oads it may be because the system pressure is too low or the pump flow is too high Contact our Customer Support 4 Additional dimensions are shown on page 20 Section II General Information Description...

Page 11: ...nit to operate at its full rated capacity If the cooling water does meet these standards the cooling capacity will be derated The chart is based on a 10 C difference between the temperature of the coo...

Page 12: ...t and are labelled FACILITYWATER and RECIRCULATING WATER FACILITY WATER connec tions are 1 inch nominal FPT RECIRCULATING WATER connections are 1 inch nominal FPT See illustration on page 20 Before in...

Page 13: ...f automotive anti freeze will void the manufacturer s warranty We recommend using distilled deionized water with a 0 05 to 0 1 MOhmcm reading Highly distilled deionized water above the 3 MOhmcm region...

Page 14: ...ng on the solenoid valve to the make up fluid source using5 16 or 3 8 inch ID flexible tubing Tubing is available from Thermo Fisher Contact our Sales Department for more information see Preface After...

Page 15: ...ted tripping of the monitor If the unit shuts down top off the reservoir and restart it When the system is full the reservoir level will no longer drop when the RECIRCULATING FLOW CONTROL valve is ope...

Page 16: ...ng temperature remember that the lowest achiev able temperature is a function of the available flow rate the temperature of the cooling water supply and the heat load The temperature control system ac...

Page 17: ...ss YES to displayCooL Press NEXT ENTER to display Stor press YES to accept all the changes pressNO to abort all changes and restore all the previous values The controller will again display the fluid...

Page 18: ...nnect the power cord from the power source and turn off the facility cooling water Place a container under the strainer to collect any water that spills out of the basket when it is removed Unscrew th...

Page 19: ...le 200 208V and 220 240V Set the switch for the appropriate range Replace the top cover Fuses The main fuses are 20 Amp RK5 The controller fuses are 1 Amp time delay Do not replace with an alternate a...

Page 20: ...s display In the event the unit is not operational the software version and the check sum can also be read from the label which is on the microprocessor chip itself The following chart uses an example...

Page 21: ...ff in the Setup Loop If Er16 is displayed the unit needs calibration contact us for assistance Display Indication Er 04 13 Interrupt error Er 14 Synchronous error Er 15 Asynchronous error Er 16 Bad ca...

Page 22: ...er sale Support Before calling please obtain the following information Part number Serialnumber Voltage of unit Voltage of power source SoftwareVersion Technical Support Our Service Department can pro...

Page 23: ...19 Section VII Diagrams Flow Diagram...

Page 24: ...20 Rear View...

Page 25: ...nized water Do not use untreated tap water as the total ionized solids level may be too high The desired level for long time usage is 1 to 3 MOhmcm compensated to 25 C The above recommendations will r...

Page 26: ...r installation or maintenance or to any unit that has been operated or maintained in any way contrary to the operating or maintenance instructions specified in this Instruction and Operation Manual Th...

Page 27: ......

Page 28: ...Thermo Fisher Scientific 81 Wyman Street P O Box 9046 Waltham Massachusetts 02454 9046 United States www thermofisher com...

Reviews: