Installation of the device
Environmental Conditions
The device must only be operated at locations that meet the particular ambient conditions listed
below:
Requirements:
• Draft-free and dry location.
• The minimal distance to adjacent surfaces must be observed on all sides, see
• The operating room must be equipped with appropriate room ventilation.
• Solid, level, fire-proof surface.
• Vibration-proof substructure (floor stand, lab table) capable of bearing the weight of the device
and of accessories (particularly if several devices are stacked).
• The electrical system of the device has been designed for an operating height of up to 2000 m
above sea level.
• To ensure a constant incubation temperature of 37 °C, the ambient temperature must be within a
range of +18 °C to +34 °C.
• Relative humidity up to 80% (max.).
• No direct exposure to sunlight.
• Devices that produce excessive heat must not be placed near the location of