Installation of the device
Retrofitting/Modifications
2-6
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
The devices are secured to one another by their own weight.
.
Retrofitting/Modifications
The following components can be retrofitted to the standard version:
S
TERI-CYCLE
i160
• threefold gas-tight screen (the gas-tight screen is installed in place of the glass door),
• door hinge replacement on outer door and glass door,
• lockable outer door,
• infrared (IR) CO
2
sensor (replaces the standard TCD sensor),
• 4-20 mA data interface.
• gas cylinder monitor
S
TERI-CYCLE
i250
• threefold gas-tight screen (the gas-tight screen is installed in place of the glass door),
• door hinge replacement on outer door and glass door,
• lockable outer door,
• split inserts,
• infrared (IR) CO
2
sensor (replaces the standard TCD sensor),
• 4-20 mA data interface.
• gas cylinder monitor
CAUTION
Transporting stacked devices!
Stacking elements are no connection elements. Stacked devices provided with a mobile base rack
must only be moved inside rooms on even floors without any slopes.
Note
Fastening on mobile racks:
If the devices are installed onto mobile racks, make sure that the rollers are secured with a locking
brake during the operation of the incubators and that the rollers are oriented toward the front for
increased stability.
Condensate formation during the operation of stacked devices:
As a rule when operating stacked devices of type
S
TERI-CYCLE
i160
and
i250
, an adapter plate should
be placed between the two devices for thermal separation.
If stacked devices are operated at an ambient temperature of more than 28 °C, an overtemperature
alarm is triggered on the upper device while the steri-run decontamination routine is run for the
lower device. Condensate formation may then occur on the upper device.
Note
Modifications:
Retrofittings and modifications must only be performed by the Technical Service of Thermo
Electron LED GmbH.
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Page 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...