10 | Operation
SpeedVac™ SPD210
Routine checks
:
1. Check all the hoses to ensure that they are secure.
2. Ensure all glass condensation flask in emptied, cleaned
and checked for crack before every run.
3. Clean the cover, cover seal and chamber with a soft lint
free cloth with one of the following:
•
Mild detergent solution
•
Diluted Methanol (50%)
•
Diluted Ethanol (50%)
Rotors and Rotor Blocks
Traditional Rotors for SPD210 Vacuum Concentrator*
*Requires purchase of adapter cat. no. RAD200.
WARNING:
Do not use this device in radioactive,
highly reactive or explosive atmosphere.
Do not use this device to process any explosive,
radioactive, highly reactive or explosive
atmosphere creating substances.
CAUTION: Product Maintenance
:
Maintenance of product only permitted to be
performed by trained service staff using suitable
measuring and auxiliary equipment as well as
detailed service instruction.
Wear gloves, eye protection, masks and lab
coats while working on the system.
WARNING: Burns due to hot device parts:
When the concentrator is set to a temperature of
60°C and above the rotor and the sample tubes
will be hot, do not touch the wall of the rotor
chamber, use suitable auxiliary to remove the
sample tubes.
Type
Description
No. of Tubes
Working Volumes (ml)
Rotor Part No
Microcentrifuge
Tubes
1.5 - 2.0 ml
200
1.2 - 1.6
RH200-12
17x60 mm (5 ml)
10
3.5
RH10-15
Glass and Plastic
Tubes
12x75 mm
200
4
RH200-12
13x100 mm
32
8
RH32-13
16x25 mm, 18x125 mm
48
12
RH48-18-125
18x150 mm, 16x125 mm
32
24
RH32-18-150
16x125 mm
88
24
RH88-16-125
Centrifuge Tubes
50 ml conical (28x115 mm)
48
40
RH48-50U
Flasks
100 ml pear-shaped flask
8
80
RH8-200
Vials
1-Dram vial (15x45 mm),
4 ml
192
3
RH192-15
20x60 mm, 20 ml
scintillation vial
50
16
RH50-28-60
Microwell Plates
Shallow well
12
-
MPTR12-210
Deep well (Shallow well)
4
-
MPTR8-210