10  |   Operation

SpeedVac™ SPD210

Routine checks

:

1. Check all the hoses to ensure that they are secure.

2. Ensure all glass condensation flask in emptied, cleaned 

and checked for crack before every run.

3. Clean the cover, cover seal and chamber with a soft lint 

free cloth with one of the following:

Mild detergent solution

Diluted Methanol (50%)

Diluted Ethanol (50%)

Rotors and Rotor Blocks

Traditional Rotors for SPD210 Vacuum Concentrator*

*Requires purchase of adapter cat. no. RAD200.

WARNING: 

Do not use this device in radioactive, 

highly reactive or explosive atmosphere.

Do not use this device to process any explosive, 
radioactive, highly reactive or explosive 
atmosphere creating substances.

CAUTION: Product Maintenance

Maintenance of product only permitted to be 
performed by trained service staff using suitable 
measuring and auxiliary equipment as well as 
detailed service instruction.

Wear gloves, eye protection, masks and lab 
coats while working on the system.

WARNING: Burns due to hot device parts:

 

When the concentrator is set to a temperature of 
60°C and above the rotor and the sample tubes 
will be hot, do not touch the wall of the rotor 
chamber, use suitable auxiliary to remove the 
sample tubes.

Type

Description

No. of Tubes

Working Volumes (ml)

Rotor Part No

Microcentrifuge 
Tubes

1.5 - 2.0 ml

200

1.2 - 1.6

RH200-12

17x60 mm (5 ml)

10

3.5

RH10-15

Glass and Plastic 
Tubes

12x75 mm

200

4

RH200-12

13x100 mm

32

8

RH32-13

16x25 mm, 18x125 mm

48

12

RH48-18-125

18x150 mm, 16x125 mm

32

24

RH32-18-150

16x125 mm

88

24

RH88-16-125

Centrifuge Tubes

50 ml conical (28x115 mm)

48

40

RH48-50U

Flasks

100 ml pear-shaped flask

8

80

RH8-200

Vials

1-Dram vial (15x45 mm), 
4 ml

192

3

RH192-15

20x60 mm, 20 ml 
scintillation vial

50

16

RH50-28-60

Microwell Plates

Shallow well

12

-

MPTR12-210

Deep well (Shallow well)

4

-

MPTR8-210

Summary of Contents for SpeedVac SPD210

Page 1: ...Installation and Operation 80302131 Revision A September 2018 SpeedVac SPD210 Vacuum Concentrator ...

Page 2: ...rmo Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited except with the written authorization of Thermo Fisher Scientific Inc The contents of this document are subject to change without notice All technical info...

Page 3: ...ion 4 Contents 4 Installation of SPD210 4 Operation 5 Description Of Control Panel 5 Manual Run 7 Auto Run 7 Drying Rate 8 Exporting Live Run Data 8 HyperTerminal Configuration 8 Lid Stay Mechanism 9 Other Topics 9 Rotors and Rotor Blocks 10 Accessories 11 Appendix 1 12 Additional SPD210 System Setup 12 Appendix 2 13 Warranty 14 ...

Page 4: ...is completed with the addition of a vapor condensation trap available separately for virtually 100 solvent recovery This results in a thorough evaporation process in a safe environment for both the sample and the user The unit features an advanced front panel with dual timer for automatic or manual control of run conditions it also features three programs that are modifiable which enable user to r...

Page 5: ...dicates extreme low temperatures and high risk of frostbite Do not touch bare metal or samples with unprotected body parts This symbol indicates possible pinch points which may cause personal injury This symbol indicates a need to use gloves during the indicated procedures If performing decontamination procedures use chemically resistant gloves Before installing using or maintaining this product p...

Page 6: ...per Installation Categories II Operative Power 115 VAC 60 Hz 8 amps 230 VAC 50 Hz 6 amps Vacuum Chamber TEFLON coated chemical resistant aluminum casting Cover Glass cover includes a safety interlock Induction Motor Maintenance free Bearings Standard Chamber temperature 35 C to 65 C 5 C increments Weight 117 lbs 53 kg Dimensions W x D x H 19 in x 26 in x 18 in 48 cm x 66 cm x 46 cm Fuse 8 A 250 VA...

Page 7: ...units configured for 230 VAC 50 Hz should be plugged into a circuit rated for at least 6 amps Note Do not use any detachable power cord that is not adequately rated for the unit Installation of SPD210 Figure 1 Rear View of System Table 1 Contents Qty Description 1 Thermo Scientific Savant SpeedVac SPD210 Concentrator 1 Cover Lock Emergency Release Tool 1 Line Cord 1 To prevent damage to pump due t...

Page 8: ...the pin with the groove Secure the assembly by screwing the retaining knob into the drive shaft above the rotor Tighten it firmly but not excessively Description Of Control Panel 1 Lid locked Indicates if the lid of the concentrator is locked or not The indicator will be illuminated when locked 2 Program buttons 3 programs that are modifiable Loading a program is achieved by pressing the correspon...

Page 9: ...parameter 9 View During a run pressing view shows the set run parameters When no run is ongoing pressing view displays the actual live parameters The time display shows either 0 00 or the end status of the last executed run e g End Err 10 Manual run Starts a Manual run based on the parameters loaded in the screen use manual run function if you do not want to set a fixed run time 11 Auto run Starts...

Page 10: ...es will click allowing air to bleed into the chamber 11 After the rotor stops spinning the cover lock disengages and the lid locked indicator is turned off The display reverts to last set parameters 12 Open the cover and remove the samples GENERAL During the run display shows actual parameters To check set parameters press View button and Select button The display will revert temporarily to the se...

Page 11: ...he data are sent at a fixed baud rate of 115200 and are comma separated as shown in the format temperature heat time run time The temperature is represented as a 2 digit integer in C The heat time is expressed as a 3 digit integer in minutes or CCC for continuous heating The runtime is expressed as a 3 digit integer in minutes Note The following section details the step by step procedures for conf...

Page 12: ...ignal the user with the Lid message on the display and an error tone Otherwise during a run the user cannot open the cover due to vacuum in the concentration chamber The cover is locked down at all times during a run and whenever power to the unit is interrupted The cover lock is an additional safety feature that reduces the risk of injury or damage while rotor is spinning Never bypass the cover l...

Page 13: ...itable measuring and auxiliary equipment as well as detailed service instruction Wear gloves eye protection masks and lab coats while working on the system WARNING Burns due to hot device parts When the concentrator is set to a temperature of 60 C and above the rotor and the sample tubes will be hot do not touch the wall of the rotor chamber use suitable auxiliary to remove the sample tubes Type D...

Page 14: ...x45 mm 1 dram 24 3 RBA24 15 45 Round or Flat Bottom Flask 2 50 ml each flask RBA2 FLSK 100 1 175 ml each flask RBA1 FLSK 250 Part No Description GF4000 Glass condensation flask 4 liter 121 6162 00 White Foam Insulating ring for coverglass trap SCC1 CryoCool heat transfer fluid 1 liter RAD200 Rotor Adapter for the SUMAX400 Upper Magnet assembly ANT100 Post trap assembly for Oligo Preps Used on UVS8...

Page 15: ...stem Figure 6 Component High Vacuum System SpeedVac Concentrator SPD210 Vacuum Tubing To Pump UVS450 ANT100 Exhaust recovery Post Trap Bleeder Valve Universal Vacuum Source SpeedVac Concentrator SPD210 Flexible Vacuum Tubing RVT450 Cold Trap SCT120 DVG50 VLP120 Vacuum Pump Chemical Trap Bleeder Valve ...

Page 16: ...es with methanol to remove materials that inhibit or prevent evaporation of aqueous samples 2 TO KEEP VACUUM PUMP OIL CLEAN TRY THE FOLLOWING Empty solvent s collected in the Refrigerated Vapor Trap as frequently as possible Change the oil in the vacuum pump when it becomes brown in color or smells of solvents being evaporated Attach a Thermo Scientific VPOF110 Vacuum Pump Oil Filter Recirculator ...

Page 17: ...01450 Power switch Molveno SX8231881100F000 1 14 80301442 MAIN CONTROL BOARD 1 15 80302069 EJOT WN1451 K40 LENGTH 8mm TORX KN6038 4 16 80302062 TORX PAN HEAD 6 32 X 3 8in 6 17 80302061 TORX PAN HEAD 6 32 X 1 4in 20 18 M50 0130 00 NUT HEX KEPS 832 SS 4 19 80302072 Torx 8 32 x 1 2 Flathead screw 4 20 80302172 TORX FLAT HEAD SCREW 6 32x5 8in 2 21 80301452 USB CABLE B MALE TO B FEMALE PANEL MOUNT 1 22...

Page 18: ...4 M60 0015 01 TUBULAR HEATER 120V 1 5 80301582 MOTOR DRIVE MAGNET ASSY 115V 1 6 80302067 TORX PAN HEAD SCREW 8 32x5 16 6 7 80302061 TORX PAN HEAD 6 32 X 1 4in 4 A B TITLE 1 2 SolidWorks 2007 10 25 C D E F E D C B A F 3 4 5 6 7 8 9 9 8 7 6 5 4 3 2 1 A B TITLE 1 2 SolidWorks 2007 10 25 C D E F E D C B A F 3 4 5 6 7 8 9 9 8 7 6 5 4 3 2 1 ASSEMBLY CHAMBER HEATER SPD210 115 ...

Page 19: ... SX8231881100F000 1 14 80301442 MAIN CONTROL BOARD 1 15 80302069 EJOT WN1451 K40 LENGTH 8mm TORX KN6038 4 16 80302066 TORX FLAT HEAD SCREW 6 32x5 16in 2 17 80302062 TORX PAN HEAD 6 32 X 3 8in 6 18 80302061 TORX PAN HEAD 6 32 X 1 4in 20 19 M50 0130 00 NUT HEX KEPS 832 SS 4 20 80302072 Torx 8 32 x 1 2 Flathead screw 4 21 80302172 TORX FLAT HEAD SCREW 6 32x5 8in 2 22 80301452 USB CABLE B MALE TO B FE...

Page 20: ... 1 4 M60 0015 02 TUBULAR HEATER 240V 1 5 80301505 MOTOR DRIVE MAGNET ASSY 1 6 80302067 TORX PAN HEAD SCREW 8 32x5 16 6 7 80302061 TORX PAN HEAD 6 32 X 1 4in 4 A B TITLE 1 2 SolidWorks 2007 10 25 C D E F E D C B A F 3 4 5 6 7 8 9 9 8 7 6 5 4 3 2 1 A B TITLE 1 2 SolidWorks 2007 10 25 C D E F E D C B A F 3 4 5 6 7 8 9 9 8 7 6 5 4 3 2 1 ASSEMBLY CHAMBER HEATER SPD210 230 ...

Page 21: ... 80301449 DAMPER 3M SJ57B1 12 7x1 3 mm 2 5 80301482 Glass cover 1 6 80302033 NUT BRACKET ASSY 2 7 80302022 HINGE SOUTHCO N6 4E 422 20 2 8 80302072 Torx 8 32 x 1 2 Flathead screw 4 9 M90 0050 01 ADHESIVE FOAM STRIP 1 16 X 1 8 x 3 inch 10 80302066 TORX FLAT HEAD SCREW 6 32x5 16in 8 11 80302082 TORX PAN HEAD 6 32 X 5 16in 2 A B TITLE 1 2 SolidWorks 2007 10 25 C D E F E D C B A F 3 4 5 6 7 8 9 9 8 7 6...

Page 22: ...MA number issued by Thermo Costs of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this manual When returning apparatus that may contain hazardous material you must pack and label them following U S Department of Transportation DOT regulations applying to transportation of hazardous materials Your shipping documents must also meet DOT regulat...

Page 23: ... 30 France 33 2 2803 2180 33 2 2803 2180 Germany 0800 1 536 376 49 6184 90 6000 0800 1 536 376 Italy 32 02 95059 552 39 02 95059 552 432 254 375 Netherlands 31 76 579 55 55 31 76 571 4440 Nordic Baltic CIS 358 9 329 10200 358 9 329 100 Russia 7 812 703 4215 7 812 703 4215 Spain Portugal 34 93 223 09 18 34 93 223 09 18 Switzerland 41 44 454 12 22 41 44 454 12 12 UK Ireland 44 870 609 9203 44 870 60...

Reviews: