Maintenance and Care
Cleaning
Thermo Scientific
Sorvall ST 8 Centrifuge
47
Clean centrifuge and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
5. Remove the centrifuge tubes and adaptors.
6. Use a neutral cleaning agent with a pH value between 6 and 8 for cleaning.
7. Dry all of the rotors and accessorie after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50 °C.
•
After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (part no.
70009824). Also treat the cavities with oil.
•
Tread the bolt of the swing out rotor with bolt grease (part no. 75003786).
CAUTION
Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
CAUTION
When cleaning, do not allow liquids, especially organic solvents, to get on the
drive shaft or the bearings of the centrifuge.
Organic solvents break down the grease in the motor bearing. The drive shaft could lock
up.
Summary of Contents for Sorvall ST 8
Page 2: ......
Page 3: ...Thermo Scientific Sorvall ST 8 Centrifuge Instruction Manual 50138912 a February 2013 ...
Page 13: ...Preface 14 Sorvall ST 8 Centrifuge Thermo Scientific ...
Page 25: ...2 Before Use Shipping the Centrifuge 26 Sorvall ST 8 Centrifuge Thermo Scientific ...
Page 43: ...5 System Menu Description 44 Sorvall ST 8 Centrifuge Thermo Scientific ...
Page 51: ...7 Disposal 52 Sorvall ST 8 Centrifuge Thermo Scientific ...
Page 65: ...10 Contact 66 Sorvall ST 8 Centrifuge Thermo Scientific ...
Page 67: ......