54 / 94
Operation
3. 5. 3. Pre-selecting the RCF-Value
1. Enter the desired value by pressing the up or down
ARROW
buttons
repeatedly,
until the desired value shows. You can adjust the RCF pre-selected value in steps of
100 x g.
2. Press
START
to confirm the pre-selected value.
If you do not press any button, the display flashes for a few seconds. The new
preselected value is now stored and the display shows the actual value.
NOTICE
The RCF value can be set to a minimum of 100 x g.
The maximum speed depends on the
centrifuge variant. The displayed RCF value is always corresponding to the maximum of centrifuge
radius of the 24 x 1.5 / 2.0 mL Rotor (75003424). Refer to
"RCF Value Explained" on page 52
for
further information.
NOTICE
Avoid if possible speed ranges close to natural resonances of the system. Runs at
resonance speeds may cause vibration and have an adverse effect on the quality of separation.
3. 5. 4. Pre-selecting Run Time
NOTICE
You can select a run time between 1 and 99 min or continuous operation.
1. Enter the desired value by pressing the up or down
ARROW
buttons
below the
display in the middle repeatedly, until the desired value shows. You can adjust the run
time in steps of 1 min.
2. Press
START
to confirm the pre-selected value.
If you do not press any key, the display flashes for a few seconds. The new preselected
value is now stored and the display shows the actual value.