Errors E01 and E02 are critical heating errors. Error Handler will lock out
heating and cooling functionality if these errors are detected. Shaking
functionality is unaffected. If the condition that caused the error is no
longer present, pressing the any button or unplugging the unit will clear
Errors.
Error E06 and E08 are non-critical errors and the unit will continue to
attempt to reach set point. If the condition that caused E06 or E08 is no
longer present, the unit will automatically clear them.
Errors E12, E13, and ‘bAL’ are shaking errors. Error Handler will lock out
shaking functions if shaking error is detected. Pressing any button or
unplugging the unit will clear the errors.
Error E21 is a data flash memory error and can be corrected by following
the steps below:
• Press any key to clear the error, this re-initializes data flash
memory.
• Note that it may be necessary to recalibrate the unit and/or re-enter
custom settings as the unit has been reset to factory default.
• The unit can now be started normally using the POWER key.