9
G
ENERAL
S
PECIFICATIONS
Catalog Number
Alarm Speed
Speed Shut Off
Timer
Unbalanced Load
Motor
Soft Start Feature
SHKA4450
SHKA4450CC
None
None
Continuous operation
(hold) or timed
operation from 1 to
60 minutes.
None
Permanent Magnet
DC
None
SHKA4450-1CE
SHKA4450CC-1CE
None
None
Continuous operation
(hold) or timed
operation from 1 to
60 minutes.
None
Permanent Magnet
DC
None
SHKE4450
SHKE4450CC
Audible with flashing
led indicate when
speed deviates more
than 10% of set point
When speed
deviates 10% of set
point unit will shut
down immediately
Beeps twice when
time has expired.
Shaking motion
stops.
If the unit is running
in an unbalanced
condition, an alarm
will sound and the
shaker will stop until
the end user corrects
the condition. The
speed display will
flash "bAL" on speed
panel LED.
Solid State Brushless
DC
Software algorithms
prevent sudden
start/stops.
SHKE4450-1CE
SHKE4450CC-1CE
Audible with flashing
led indicate when
speed deviates more
than 10% of set point
When speed
deviates 10% of set
point unit will shut
down immediately
Beeps twice when
time has expired.
Shaking motion
stops.
If the unit is running
in an unbalanced
condition, an alarm
will sound and the
shaker will stop until
the end user corrects
the condition. The
speed display will
flash "bAL" on speed
panel LED.
Solid State Brushless
DC
Software algorithms
prevent sudden