Adjusting the Gain Settings
Gain Control Settings
Thermo Fisher Scientific
SGD-O User Guide
7-3
Table 7–1.
FSU setup to obtain required voltage reading
Pipe Size
Source Size
C13
(Default 330 pF)
Gain Setting
(Default 1X)
2” HP
100 mCi
330 or 470 pF
1X or 2X
2” HP
145 mCi
330 or 470 pF
1X
2” HP
200 mCi
330, 470, or 680 pF
1X
3” HP
200 mCi
330 pF
1X
4” HP LP
200 mCi
330 or 470 pF
1X or 2X
6” LP
200 mCi
330 or 470 pF
2X
8” LP
200 mCi
220 or 330 pF
2X
10” LP
200 mCi
220, 330, or 470 pF
2X or 3X
Figure 7–2 shows how the jumpers and resistors work together to control
the gain of the AR1 op-amp.
Figure 7–2.
Gain control of the AR1 op-amp
Gain Control
Settings