6-1
Chapter 6
The Current Output
The standard 4–20 mA output can be adjusted so the control room and
local display readings are identical. In order to do this and to check the
connection, hold commands will be used to set the current output to
known value.
Note
for instructions on how to use a DAC.
1.
Hold the current output at 4 mA (DAC
8
).
If the value is not 4 mA after executing this command, you can apply a
correction factor. First, determine the correction factor by dividing the
new value by the initial value. For example, if the current output is
3.98 mA, divide 4.0 by 3.98:
4.0/3.98 = 1.005.
In this case, the correction factor is 1.005. Enter the correction factor
(DAC
053013
).
Note
The initial value in this formula is the measured value of the 4 mA
output when the correction factor is equal to 1.0.
2.
Hold the current output at 20 mA (DAC
5
).
If the value is not 20 mA after executing this command, you can apply
a correction factor. First, determine the correction factor by dividing
the new value by the initial value. For example, if the current output is
20.04 mA, divide 20.0 by 20.04:
20.0/20.04 = 0.998.
In this case, the correction factor is 0.998. Enter the correction factor
(DAC
052013
).
Note
The initial value in this formula is the measured value of the 20 mA
output when the correction factor is equal to 1.0.
3.
Hold the current output at 12 mA (DAC
2
).
Hold the Current
Output
Summary of Contents for SGD-O
Page 2: ......
Page 3: ...SGD O Non Contacting Density Gauge For Fracturing Cementing User Guide P N 717910 Revision D...
Page 4: ......
Page 6: ...This page intentionally left blank...
Page 8: ...This page intentionally left blank...
Page 36: ...This page intentionally left blank...
Page 66: ...This page intentionally left blank...
Page 70: ...This page intentionally left blank...