Redox/ORP Electrode User Guide
Electrode Calibration
ORP readings can be reported using the absolute mV values
or the E
H
mV values. E
H
mV values give the corresponding
mV values that would be obtained using a Standard Hydrogen
Electrode (SHE), also known as the Normal Hydrogen Electrode
(NHE). Orion ORP standard, Cat. No. 967901 or 967961, allows
operators to correlate measurements to the E
H
mV value with
an Orion ORP electrode.
The relative mV mode can be used to calibrate the ORP
electrode so the electrode reads the E
H
mV values in samples.
When an ORP electrode is calibrated to read E
H
mV values, the
resulting sample readings can be compared among multiple
meter and electrode systems.
Note:
If the meter can perform an automatic ORP calibration
using an E
H
mode, refer to the meter user guide for detailed
instructions on the meter calibration procedure.
1. Set the meter to the relative millivolt mode. Refer to the
meter user guide for detailed instructions.
2. Place the ORP electrode in ORP standard, Cat. No. 967901
or 967961.
3. Set the meter to the E
H
mV value of the ORP standard at
the measured temperature using the information in
Table 1
.
Table 1– ORP Standard Values
Absolute mV values may vary by ± 60 mV
Temperature
(°C)
E
H
Value
(mV)
Absolute Value
with Cat. No.
900011 Filling
Solution (mV)
Absolute Value
with Cat. No.
900001 Filling
Solution (mV)
0
438
218
176
1
437
218
176
2
437
218
176
3
436
218
176
4
435
218
176
5
435
218
176
6
434
218
176
7
433
218
176
8
433
218
175
9
432
219
175