Redox/ORP Electrode User Guide

6.  Calculate the normality of the sample, N

x

, in equivalents per 

liter.

 

N

x

  =  (V

t

 * N

t

) / V

x

where:

 

N

t

  =  normality of titrant (Eq/1) 

 

V

t

  =  volume of titrant at end point (mL)

 

V

x

  =  volume of sample (mL)

Table 2 References:

1.  Furman (ed.), Scott’s Standard Methods of Chemical 

Analysis, 5th Ed., Van Nostrand, Princeton, N.J. 1938.

2.  Hillebrand, Lundell, Bright, and Hoffman, Applied Inorganic 

Analysis, 2nd Ed., Wiley, New York, 1953.

3.  Kolthoff and Sandell, Testbook of Quantitative Inorganic 

Analysis, 3rd. Ed., MacMillan, New York, 1952.

4.  Oesper, Newer Methods of Volumetric Chemical Analysis, 

Van Nostrand, Princeton, N.J., 1938.

5.  Treadwell and Hall, “Analytical Chemistry,” vol. 2, 

Quantitative Analysis, 8th Ed., Wiley, New York, 1935.

Summary of Contents for REDOX/ORP 9678BNWP

Page 1: ...er Scientific l Instruments is Instruments Center 915 USA 000 25 1480 232 6015 232 6031 om water User Guide Redox ORP Electrodes...

Page 2: ......

Page 3: ...rugged environments Glass Combination Redox ORP Electrodes Cat No 9778BNWP and 9778SC The glass body enables the electrode to be used in solutions that contain organic solvents Use the electrode in wa...

Page 4: ...lectrodes have a temperature connector that is compatible with the Star Series meters The 9180BN and 9179BN ORP ATC Triode electrode have a temperature connector that is compatible with the A Series m...

Page 5: ...RP electrodes contain a silver silver chloride Ag AgCl reference that may become clogged in solutions that contain silver complexing or binding agents such as TRIS proteins and sulfides Frequent clean...

Page 6: ...uld be open whenever the electrode is in use 4 For Sure Flow electrode only wet the junction by pressing down on the electrode cap to let a few drops of the filling solution flow out of the electrode...

Page 7: ...erence reference filling solution chamber Sure Flow junction sensing element Figure 1 Refillable ORP Electrode filling hole Ag AgCl internal reference filling solution chamber ceramic reference juncti...

Page 8: ...ed among multiple meter and electrode systems Note If the meter can perform an automatic ORP calibration using an EH mode refer to the meter user guide for detailed instructions on the meter calibrati...

Page 9: ...25 219 174 19 424 219 174 20 424 219 174 21 423 219 174 22 422 219 174 23 421 219 173 24 420 220 173 25 420 220 173 26 419 220 173 27 418 220 173 28 417 220 172 29 416 220 172 30 415 220 172 31 414 22...

Page 10: ...ents If the electrode is used in dirty or viscous samples or the electrode response becomes sluggish empty the electrode completely and hold the junction under running water Empty any water from the e...

Page 11: ...nd other water samples 1 Prepare the electrode according to the Electrode Preparation section 2 Set the meter to the millivolt relative millivolt or EH measurement mode 3 Place the electrode into the...

Page 12: ...nts Table 2 lists suitable titrants for different unknowns the equation for the titration reaction interferences and a reference where information about preparing the sample and running the titration...

Page 13: ...andard Methods of Chemical Analysis 5th Ed Van Nostrand Princeton N J 1938 2 Hillebrand Lundell Bright and Hoffman Applied Inorganic Analysis 2nd Ed Wiley New York 1953 3 Kolthoff and Sandell Testbook...

Page 14: ...ion and cause false high results iron as Fe 2 Fe 3 or in steel high levels of organic acids such as acetic oxalic tartaric and alcohols F if present in more than trace amounts Cl PO4 3 remove by addin...

Page 15: ...2 2I S4O6 2 3 ceric sulfate Ce SO4 2 not recommended for samples containing Cl Fe 2 Ce 4 Ce 3 Fe 3 2 potassium dichromate K2Cr3O7 recommended for samples containing Cl Cr2O7 2 6Fe 2 14H 2Cr 3 6Fe 3 7H...

Page 16: ...e is refillable drain the reference chamber flush it with fresh filling solution and refill the chamber with fresh filling solution 4 If the electrode is refillable make sure that the filling solution...

Page 17: ...fast rate 2 If the electrode is refillable drain the reference chamber and refill it with fresh filling solution 3 Thoroughly rinse the electrode with distilled water Cleaning Solutions Cat No 900021...

Page 18: ...ode and then refill any lost filling solution Disassembling the Sure Flow Electrode Note Disassembly is usually not required and should not be done unless a thorough cleaning is required 1 Tip the ele...

Page 19: ...e Storage Short term Storage up to one day To ensure a quick response and free flowing junction the reference junction must not dry out It is recommended that the electrode be soaked in a 4 M potassiu...

Page 20: ...he reading stabilizes 3 The mV potential should be 220 mV 60 mV If the mV reading is out of range repeat the measurement with fresh ORP standard If the electrode fails this procedure clean the electro...

Page 21: ...3 39 g reagent grade KF 2H2O Place in a 100 mL volumetric flask Add 50 mL distilled water and swirl to dissolve solids Dilute to volume with distilled water 3 Transfer solution A to a 150 mL beaker Pl...