PRO902C MANUAL
T
FS
Rev 6.0
August 16, 2007
1-2
7. Heated filter control module
8. Valve control module (Watchdog)
9. Valve manifold
Tubing and wiring umbilical (please refer to the 1200 series Umbilical Drawings):
1. Tubing umbilical
2. Wiring umbilicals - paired wiring umbilical, 12 shielded pairs, #20
AWG in PVC jacket.
Remote probe controller assembly: please refer to the 7000 series probe controller
drawings.
1.3
Dilution probe conditioning assembly hardware
The PRO902C dilution probe conditioning assembly consists of eight subassemblies used to
condition, analyze, and transport the sample:
1. Probe enclosure.
2. Enclosure heater.
3. Valve manifold and dilution air regulator.
4. Heated filter, dilution eductor, and probe barrel assembly.
5. Heat exchanger and drain valve assembly.
6. TEI O
2
analysis system.
7. Probe power, purge and temperature control.
8. Absolute pressure transducer.
1.3.1 Probe enclosure
The dilution probe and conditioning assembly is housed in a fiberglass enclosure that
measures approximately 24 inches (H), by 24 inches (W), by 9.75 inches (D). The enclosure
is designed to protect the instruments and electrical controls from highly corrosive
atmospheres.
1.3.2 Enclosure heater
The enclosure heater is located on the bottom, right-hand side of the enclosure (please refer
to the 7000 series probe box assembly drawing, item 17). The assembly consists of one 400-
watt heat element, a 30 CFM fan, and an adjustable thermostat switch. The fan motor runs
continuously when power is supplied to the probe conditioning enclosure heater. The heat
element cycles on and off according to the temperature inside the enclosure. The temperature
of the enclosure is maintained at 100°F by the adjustable thermostat.