Chapter 7
|
Data Transfer and Software Updates
Command
Command Details
SETMODE
SETMODE MODE <CR>
Set the current channel measurement MODE as:
PH
COND
DOSAT
RMV
RES
DOCON
ORP
SALT
ISE
TDS
Example: SETMODE PH <CR>
GETMODE
GETMODE CHANNEL <CR>
Print the channel measurement mode
Example: GETMODE 1
SETCSV
SETCSV <CR>
Set printing format to CSV
SETKEYLOCK
SETKEYLOCK NUMBER <CR>
To disable the key pad: Number = 0
To enable the key pad Number = 1
Meter Measurement Data from GETMEAS Command
Measure
Mode
PC (CSV) Format
pH
Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,
Channel, Mode, pH Value, pH Unit, mV Value, mV Unit, Temperature Value, Temperature
Unit, Slope Value, Slope Unit, Method #, Log #
Example:
A211 pH, X01036, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, pH, 7.000, pH, 0.0,
mV, 25.0, C, 98.1,%, M100, #1 <CR>
mV
Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,
Channel, Mode, mV Value, mV Unit, Temperature Value, Temperature Unit, Method #,
Log #
Example:
A211 pH, X01036, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, mV, 0.0, mV, 25.0,
C, M100, #2 <CR>
RmV
Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,
Channel, Mode, Relative mV Value, Relative mV Unit, Offset mV Value, Offset mV Unit,
Temperature Value, Temperature Unit, Method #, Log #
Example:
A211 pH, X01036, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, RmV, 0.0, RmV,
0.0, mV, 25.0, C, M100, #3 <CR>
ORP
Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,
Channel, Mode, ORP Value, ORP Unit, mV Value, mV Unit, Temperature Value,
Temperature Unit, Method #, Log #
Example:
A211 pH, X01036, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, ORP, 0.0, mV, 0.0,
mV, 25.0, C, M100, #4 <CR>
Thermo Scientific
Orion Star A220 Series Portable Meter User Manual
|
69