Chapter 7
|
Data Transfer and Software Updates
Printer Compatibility and Requirements
Orion Star A220 series portable meters can print directly to the Orion Star series inkjet printer,
Catalog Number 1010006. Measurement and calibration data sent from the meter to the Orion
Star series inkjet printer is automatically formatted to fit the paper width when the printer data
format is selected within the Export Data setting. The Orion Star series inkjet printer has a 9600
baud rate and it is packaged with the RS232 printer cable required to interface an Orion Star
A220 series meter to this printer.
The baud rate of Orion Star A220 series meters can be set to 1200, 2400, 4800, 9600, 19200
or 38400 for communication to different printers. The fixed meter settings are:
Number of Data Bits: 8
Stop Bits:
1
Parity:
None
Flow Control:
XON/XOFF
The meter will send measurements and calibration data to the printer if the printer option is
turned on in the setup menu. Alternatively, measurements and calibration data can be sent to
the data log and calibration log. From the data log and calibration log, the operator can chose to
print a single data log point, a range of data log points or the entire data log.
58
|
Orion Star A220 Series Portable Meter User Manual
Thermo Scientific