Chapter 3
|
Meter Setup Menus
Calibration Log
Orion Star A220 series meters save up to ten of the most recent calibrations per channel.
1.
In the measurement mode, press the setup key.
2.
Press the
p
,
q
,
t
or
u
key to highlight View Log and press the f3 (select) key.
3.
Press the
p
or
q
key to highlight Calibration Log and press the f2 (select) key.
4.
Press the
p
or
q
key to highlight pH–Channel, Conductivity–Channel or DO/RDO–
Channel and press the f2 (select) key.
a.
If pH–Channel is selected, press the
p
or
q
key to highlight pH, RmV or ORP and
press the f2 (select) key.
b.
If Conductivity–Channel is selected, press the
p
or
q
key to highlight Conductivity,
Resistivity, TDS or Salinity and press the f2 (select) key.
c.
If DO/RDO–Channel is selected, press the
p
or
q
key to highlight DO or RDO and
press the f2 (select) key.
5.
The meter will display a list of calibrations for the selected channel and calibration mode.
The list chows the sequential number of the calibration log and the data and time the
calibration was completed.
6.
To view detailed information for an individual calibration, press the
p
or
q
key to highlight
a calibration and then press the f2 (select) key.
a.
To export the calibration to a printer or computer, press the f2 (print) key.
Exported Calibration Log Example,
Printer Data Format:
Thermo Scientific
Orion Star A220 Series Portable Meter User Manual
|
33