Orion Star A215 Benchtop pH/Conductivity Meters

6

Measurement

The Orion Star A215 meter is capable of showing two measurements simultaneously on the display. The 
first channel can be set to measure pH, mV, RmV (relative mV) or ORP. The second channel can be set to 
measure Cond (conductivity), TDS, Salinity or Res (resistivity). 

The meter can be set to display one or both of these measurement channels, depending on the operator’s 
needs. In the measurement mode, press 

f3 (channel)

 to scroll through a single measurement display of 

channel one or two or a dual measurement display of channel one and two. To change the measurement 
mode of a channel, press 

f3 (channel)

 until the single measurement display of that channel is shown 

and then press 

 until the correct mode is shown.

Note: 

It is highly recommended that any unused channels not be shown on the meter display while 

taking measurements, since the meter waits for all displayed channels to stabilize before logging the 
measurement data.

Press 

 while taking a measurement in the continuous measurement mode to freeze the display and 

press 

  a second time to unfreeze the display and continue the measurement. Press 

 while taking 

a measurement to manually export the measurement  to the data log, if the data log is enabled in the 
setup menu.

1.  Rinse the pH electrode, conductivity cell and any other electrodes in use with distilled water, blot dry with 

a lint-free tissue and place into the sample.

2.  If the stirrer probe is in use, press 

 to turn on the stirrer probe.

3.  Start the measurement and wait for it to stabilize. 

a.  If the meter is in 

AUTO-READ

 mode (default setting), press 

 to start the measurement. When 

the 

 icon stops flashing, record the applicable measurement parameters and temperature of the 

sample. Press 

 again to start a new measurement.

b.  If the meter is in continuous mode, the meter will immediately start taking a measurement and 

update the display whenever the measurement changes. Wait for the display to show 

ready

 and 

record the applicable measurement parameters and temperature of the sample.

c.  If the meter is in timed mode, the meter will log measurements at the preselected time interval, 

regardless of the measurement stability. The meter will update the display whenever the 
measurement changes, so the applicable measurement parameters and temperature of the sample 
can be recorded when the display shows 

ready

.

4.  If the stirrer probe is in use, press 

 to turn off the stirrer probe.

5.  Remove the electrodes from the sample, rinse with distilled water, blot dry and place into the next sample.

6.  Repeat steps 2 through 5 for all samples.

7.  When all samples have been measured, store the electrodes according to their user guides.

Summary of Contents for Orion Star A215

Page 1: ...ding to the directions in the electrode user guide 2 Connect the appropriate items as labeled on the meter and as shown in the gure to the right Electrode Arm The electrode arm can be attached to eith...

Page 2: ...t stability M 100 Indicates a method is in use and the number of the method being used pH Indicates the type of measurement and determines the type of calibration that will be performed ready Speci es...

Page 3: ...n When the meter is on press and quickly release to turn the display backlight on or off or hold down to turn the meter off In the measurement mode press to take a measurement In the setup calibration...

Page 4: ...t to accept the displayed value b Press f3 edit to access the numeric entry screen and edit the value i Press or to highlight a number decimal point or negative sign press f3 enter to select the highl...

Page 5: ...and place into the standard 3 When the conductivity cell and standard are ready press f3 start 4 Wait for the conductivity value on the meter to stabilize and stop ashing and perform one of the follow...

Page 6: ...data log is enabled in the setup menu 1 Rinse the pH electrode conductivity cell and any other electrodes in use with distilled water blot dry with a lint free tissue and place into the sample 2 If th...

Page 7: ...bmenus a To select a value from a list of options press or to highlight the desired value and press f3 select to set the value b To enter a numeric value use the numeric entry screen i Select the valu...

Page 8: ...t Temperature Calibration t Temperature Input Method Mode Settings t Measure Mode t Read Type t Cell K t Stability t Averaging t Ref Temp t Temp Comp t Temp Coeff t Alarm Settings Temperature t Manua...

Page 9: ...usethe numericentryscreentochangethevaluesandpressf1 back whenthetimeintervaliscorrect 6 Press to return to the measurement mode pH Buffer Group Selection The selected buffer group allows for the auto...

Page 10: ...tycell a Press or to highlight a number or decimal point press f3 enter to select the highlighted item and repeat until the nominal cell constant value is shown b Press f2 done to exit the numeric ent...

Page 11: ...or EP and press f3 select 6 Press to return to the measurement mode Viewing the Calibration Log 1 In the measurement mode press 2 Press or to highlight Calibration Log and press f2 accept 3 Press or...

Page 12: ...Tel 81 045 453 9175 wai asia thermo sher com Australia Tel 613 9757 4300 in Australia 1300 735 296 InfoWaterAU thermo sher com 9 0 0 1 2 0 0 8 68X576505 RevA 0811 Viewing the Data Log 1 In the measure...

Reviews: