Orion Star A215 Benchtop pH/Conductivity Meters
10
Conductivity Nominal Cell Constant Entry
The nominal cell constant value is used during an automatic conductivity calibration and allows the meter to
determine which Thermo Scientific Orion conductivity standard is being used for the calibration.
1. In the measurement mode, press
.
2. Press
or
to highlight
Mode and Settings
and press
f3 (select)
.
3. Press
or
to highlight
Cell K
and press
f3 (select)
.
4. Press
f3 (select)
to access the numeric entry screen and enter the nominal cell constant of the conductivity cell.
a. Press
,
,
or
to highlight a number or decimal point, press
f3 (enter)
to
select the highlighted item and repeat until the nominal cell constant value is shown.
b. Press
f2 (done)
to exit the numeric entry screen.
5. Press
f1 (back)
and then press
to return to the measurement mode.
Conductivity Reference Temperature Selection
The reference temperature is the temperature that all conductivity measurements will be reported at by the
meter if temperature compensation is enabled.
1. In the measurement mode, press
.
2. Press
or
to highlight
Mode and Settings
and press
f3 (select)
.
3. Press
or
to highlight
Ref Temp.
and press
f3 (select)
.
4. Press
or
to highlight
5 °C, 10 °C. 15 °C, 20 °C
or
25 °C
and press
f3 (select)
.
5. Press
to return to the measurement mode.