Chapter 3
|
Meter Setup Menus
Electrode Stability Test Procedure
1.
In the measurement mode, press the setup key.
2.
Press the
p
,
q
,
t
or
u
key to highlight Diagnostics and press the f3 (select) key.
3.
Press the
p
or
q
key to highlight Stability Test and press the f2 (select) key.
4.
Press the
p
or
q
key to highlight pH–Channel, pH/ISE–Channel, Conductivity–Channel or
DO/RDO–Channel and press the f2 (next) key.
5.
Make sure the electrode or sensor to be tested is connected to the channel selected in the
previous step and press the f2 (yes) key.
6.
Place the electrode or sensor in an appropriate solution with a stable temperature and press
the f2 (yes) key.
a.
Thermo Scientific Orion pH 7.00 buffer (Catalog Number 910107, 910760 or 910710) is
recommended for testing a pH electrode.
b.
Thermo Scientific Orion ORP standard (Catalog Number 967901 or 967961) is
recommended for testing an ORP electrode.
c.
A calibration standard with a moderate concentration of the ion of interest is
recommended for testing an ion selective electrode (ISE).
d.
Thermo Scientific Orion 1413 µS conductivity standard (Catalog Number 011007 or
01100710) is recommended for testing most conductivity sensors. Thermo Scientific
Orion 100 µS conductivity standard (Catalog Number 011008) is recommended for
testing ultra pure water conductivity sensors.
e.
A prepared water-saturated air calibration sleeve is recommended for testing RDO
optical and polarographic dissolved oxygen sensors.
7.
The meter will perform the stability test and show the raw input reading per minute drift and
noise on the display.
8.
The meter will display Stability Test Passed or Stability Test Failed. Press the f1 (esc) key
to return to the measurement mode.
Thermo Scientific
Orion Star A210 Series Benchtop Meter User Manual
|
49
Summary of Contents for Orion Star A210 Series
Page 5: ......