Gas Connections
To connect the gas lines to the back of the Orbitrap Fusion Series MS, follow the next
procedure. Make sure that the ultra-high-purity (UHP) and high-purity (HP) supply gases are
delivered with the necessary pressure and purity as listed in
.
Table 2.
Summary of the required gases for the Orbitrap Fusion Series MS
Type
Recommended purity
Required
pressure
Typical daily
consumption
a
a
Approximate value when operating 24 hours and 7 days a week
Function
Helium
UHP (99.999%)—Has less than
1.0 ppm each of water, oxygen,
and total hydrocarbons.
275 ±70 kPa
(40 ±10 psi)
9.22 L
(0.325 ft
3
)
Linear ion trap damping gas
and collision gas
Nitrogen
UHP (99.999%)—Has less than
3.0 ppm each of water, oxygen,
and total hydrocarbons.
345 ±70 kPa
(50 ±10 psi)
46 L (1.62 ft
3
)
Venting gas, ion routing
multipole (IRM) collision
gas, and make-up gas and
reagent carrier gas for the IC
and ETD configurations
HP (99%)
690 ±140 kPa
(100 ±20 psi)
11 500–26 700 L
(406–943 ft
3
)
API source auxiliary gas,
sheath gas, and sweep gas
Note
Connections and gas delivery systems might vary. You are responsible for supplying
any additional fittings or connections necessary during installation. If your system
includes additional devices that require their own gas connections, refer to those
instruments’ manuals.
CAUTION
You must use either copper (provided) or stainless steel for the UHP gas lines.
CAUTION
After completing these connections, route the tubing so that they are not a trip
hazard.