Get Started with Your Nicolet Summit Spectrometer
Before Your Summit Spectrometer Arrives
Prepare a suitable workspace for your spectrometer before it arrives to ensure the most accurate data
and the best long-term performance.
This article explains what to do when your spectrometer arrives and introduces environmental and
electrical factors that can influence your spectrometer’s performance. For a more comprehensive
explanation of these requirements, see the Site and Safety Information for the spectrometer.
Unpacking the Spectrometer
Before you open the box, there are two important steps to take as soon as the instrument arrives:
l
Check the exterior of the shipping box for damage
If you notice any signs of damage, contact us or your local distributor for instructions.
l
Allow the spectrometer to come to room temperature
Inside the shipping box, the spectrometer is sealed in a plastic bag to keep it dry. You must allow 24
hours for the spectrometer to reach room temperature before you open the bag. If you open the
bag before the spectrometer is warmed up, condensation could form that could damage internal
optical components and cause permanent damage.
The warranty will not cover damage due to improper moving techniques or removing the sealed plastic
bag before the instrument has come to room temperature.
Preparing the Workspace
Before your instrument arrives, make sure your workspace will be able to properly support the
spectrometer. In addition to allowing enough room for the spectrometer, consider several environmental
and electrical requirements.
Spectrometer Dimensions
The Summit spectrometer requires a fairly small space, but make sure you also leave room around the
instrument so that heat can dissipate from the vents and so that you can easily access the instrument’s
ports, power switch, and cables.
l
Summit instrument weight: 9.6 Kg (21 lbs)
l
Dimensions (W x H x D): 34 cm x 24 cm x 32 cm; (13.3 in. x 9.6 in. x 12.7 in.)
6
Nicolet Summit LITE FTIR Spectrometer User Guide
Thermo Scientific