1.
—
Open the shutoff valve.
2.
—
Adjust the pressure regulator until the gauge indicates that the pressure is between 0.7
and 1.4 bar (70 to 140 kPa, or 10 to 20 psig).
Notice: Flow rates greater than 10 scfh can cause vibration that can affect data quality. We
recommend keeping the flow rate at approximately 10 scfh.
3.
—
Set the flowmeter to 10 scfh.
10 Using Your Nicolet iZ10 Module