3 - 2
Merlin
Chapter 3
Thermo Scientific
A male 9-pin D-connector is located on the rear of the chiller. The
connector is used for either 12 - 24VAC or 12 - 24VDC remote start/
stop; and as a status relay dry contact.
To start the chiller from a remote location apply the correct voltage to
pins 6 & 8. When the chiller is started by a remote start/stop voltage
on the controller will flash.
Once started remotely, stop the chiller by removing the applied voltage
from pins 6 & 8 or by pressing
on the local controller. To restart
remotely after stopping locally, remove and reapply the remote start
voltage to J101.
NOTE
Enabling serial communications disables the remote start/stop
feature.
Should the chiller experience a high/low temperature warning/fault, a
reservoir level warning/fault, or a low pump flow warning/fault, the
status relay will de-energize.
The pin out information is:
Pin #
Function
Pin #
Function
1
No Connection
6
Remote On +
2
Status, NO
7
RTD2B
3
Status, NC
8
Remote On -
4
RTD2A
9
No Connection
5
Status, Common
Hardware Mating Connector AMP Part# 745491-2 or equivalent
12 to 24 Volts DC ±10% maximum input current, 9ma.
12 to 24 Volts AC ±10% maximum input current, 5ma.
40 volts rms. continuous.
56 volts peak.
Rating for both normally open and normally closed status relay contacts
are: Maximum rated load - 1.0 Amp, 24 Volts AC/DC, resistive. Minimum
permissible load - 1 ma @ 5 VDC.
Remote Start/Stop
Remote Start Input
Absolute Maximum
Input Voltages
Summary of Contents for Neslab Merlin M 75
Page 2: ......
Page 24: ...Merlin MD EG PG pH DI EG PG DI Polar SDS 4 0 gpm 15 1 lpm Thermo Scientific Polar 19...
Page 30: ...EG PG pH pH DI EG PG DI Polar SDS 4 0 gpm 15 1 lpm Thermo Scientific Polar 19 mm...
Page 66: ......
Page 82: ...2 12 Merlin Chapter 2 Thermo Scientific...
Page 98: ...5 4 Merlin Chapter 5 ThermoScientific...
Page 103: ......