Section 12- Protein Bradford
Standard Curve Features
Standard curves can be saved and reloaded for reference use by using the ‘Standard Curve’ pull down menu and choosing save or
reload functions. Selecting the ‘View Standard Curve’ button allows the user to review the standard curve at any time.
The Delete Point
button allows the user to delete points by first selecting the data point to be deleted in the table and then choosing the
‘Delete Point’ button. If erroneous or non-representative readings are encountered for a specific standard, all replicates of that standard
are cleared by selecting ‘Reset this std’. Additionally, all standards can be deleted at once using the ‘Reset This Window’ (F11) button.
Regular Bradford curve covers 200-
8000 ug/ml. Note the linear range
is 100-1000 ug/ml
A mini-Bradford assay covers an
approximate range of 15-100 ug/ml.
Exiting the Bradford Module
It is recommended that you process all of the unknowns to be assayed before exciting the Bradford software module.