Wiring
Power Supply
Thermo Fisher Scientific
LevelPRO Installation Guide
4-5
The system is designed to operate on 24 Vdc. If the optional AC power
board is installed, the system can also operate on 115 or 230 Vac (± 15%),
50/60 Hz. The maximum input power requirement is 12 VA. If both AC
and DC power are supplied to the detector, the detector will draw power
from whichever source provides the higher DC voltage.
The maximum length of the cable that can be used to supply power varies
depending on the wire gauge. Refer to the
cable information and other wiring details.
Warning
All wiring must be done by qualified individuals in accordance
with applicable codes such as the NEC (National Electric Code) ANSI /
NFPA 70 specifications or the Canadian Electrical Code Part 1. For
intrinsically safe systems, refer to ANSI / ISA RP 12.6.
Note
To meet the requirements of CSA 1010.1, an external switch or
circuit breaker must be installed to allow the power source to be
disconnected from the gauge. In addition, protective bonding (grounding)
must always be provided, even if DC power is used.
J12 on the CPU board is the connector for the DC source voltage wiring.
connections.
Table 4–1.
DC power connections
Power Supply
J12 on CPU Board
Positive (+)
J12-2
Negative (-)
J12-1
Note
To meet the requirements of CSA 1010.1, the input DC terminals
must be supplied from a SELV (Safety Extra Low Voltage) source.
Power Supply
DC Power
Summary of Contents for LevelPRO Series
Page 2: ......
Page 3: ...LevelPRO Continuous Gamma Level System Installation Guide P N 717760 Revision H ...
Page 4: ......
Page 6: ...This page intentionally left blank ...
Page 8: ...This page intentionally left blank ...
Page 14: ...This page intentionally left blank ...
Page 28: ...This page intentionally left blank ...
Page 40: ...This page intentionally left blank ...
Page 48: ...This page intentionally left blank ...
Page 96: ...This page intentionally left blank ...
Page 106: ...This page intentionally left blank ...