background image

Thermo Scientific Jewett Blood Bank Refrigerators

Installation and Operation

1

Introduction

This product is a complete packaged unit ready to operate when 
plugged into an electrical source.  It is not necessary to have a 
refrigeration serviceman set the unit in operation.  The unit has 
been tested prior to shipment.  Read all the instructions before 
proceeding with installation.. 

2

Inspection for Damage

Uncrate the product and remove all packing and crating 
materials.  Inspect the product and parts thoroughly for damage 
or missing parts.  File all claims for damage with the 
transportation company immediately.  Do not file claims with the 
manufacturer.

3

Installation Instruction

1. Measure all doorways and passageways for clearance before 

moving cabinet.  Model sizes 17 and 37 will pass through a 
standard 30" X 84" door opening.  Model sizes 25 and 55 will 
pass through a 36" X 84" door opening.  For additional front 
to back clearance, remove the door handles using a 
screwdriver.  If additional clearance is needed, remove the 

condensate evaporator assembly, channel guards and drain 
tube from the back of the cabinet.  Remove (2) 1/4" bolts 
using a 7/16"wrench.  A 3/8" nut driver will enable removal 

of the condensate heater.  To remove the drain tube, detach 
the rubber tube located below the unit cooler inside the 
cabinet.  With a slight twisting motion, pull drain tube from 
the back of the cabinet.  Do not disconnect wiring on 
condensate evaporator assembly.  Flexible armored cable will 
permit sufficient movement to swing assembly to top or side 
of cabinet.

2. Inspect the interior, exterior and mechanical equipment for 

special instruction tags fastened at various points.

3. Move the refrigerator into the desired location.  Make sure 

the bottom of the refrigerator is evenly supported.  Thin 
shims under the points of rest can be used to equalize the 
distribution of weight.  If the cabinet sets on an uneven floor, 
a slight rocking or vibration might result when the 
condensing unit is set in operation.

4. A minimum of 8" clearance is required between the ceiling 

and the top of the cabinet; and a minimum of 3" clearance is 
required at the rear to allow adequate ventilation of the 

mechanical equipment.

5. The condensing unit is shipped with all service valves open 

and ready for operation.  Do not adjust the refrigerant valves 
or the temperature control.

6. Use of electrical characteristics, other than those specified on 

the serial plate will cause permanent damage to the 
mechanism.  The serial plate is located inside the refrigerator 
at the top of the right hand wall.

7. All models have been provided with an electric condensate 

evaporator located on the back of the cabinet.  Periodic 
cleaning of the condensate pan will insure proper operation. 

CAUTION!

PAN AND HEATER ARE HOT.

8. Operate the refrigerator for several hours to allow the unit to 

reach normal operating temperatures before storing product.  

NOTE:  Erratic operation on initial start-up does not indicate 
a faulty control.  When normal operating temperature is 
reached, the condensing unit will cycle regularly.

4

Maintenance Instructions

1. Frequent cleaning of the interior and exterior with water and 

a good fungicidal detergent that eliminates harmful bacteria, 

stains and other foreign matter will keep the unit fresh and 
new looking.

2. Shelves and/or drawers should be removed from the cabinet 

and thoroughly scrubbed. Clean door gasket periodically.

3. The unit cooler fan operates continuously when the door is 

closed and requires no lubrication.  These models require no 
manual defrosting.  The cooling coil automatically defrosts 
when the condensing unit is on the off cycle.

4. The condensing unit needs no oil or other lubrication.  The 

finned condenser can become clogged with lint or dust.  The 
openings between the fins should be kept clean.  A vacuum 
cleaner or small test tube brush works well for this purpose.  

This should be accomplished on an annual basis.  Failure to 
keep the condenser fins free of dirt and lint will result in 
erratic operation and may damage the refrigeration system.

5. Annual inspection of the mechanical refrigeration equipment 

by a qualified serviceman is recommended.  A qualified 
mechanic can frequently make adjustments that will prevent 
future breakdown.

6. The refrigeration system is charged with refrigerant R134a.  

If the system is opened for any reason, extreme care should 

be taken to prevent the entry of moisture-bearing air.  A new 
drier should be installed in the lines when the system is 
closed.

5

Operating Instructions

JEWETT Blood Bank Refrigerators are factory preset to operate 
within a +2° C to +4° C temperature range.  Manual defrosting of 
the refrigerator cabinet is not required, as the unit cooler defrosts 
automatically when the condensing unit is on the off cycle.  The 
unit cooler fan motor inside of the refrigerator runs continuously 

when the door is closed to provide a uniform cabinet 
temperature.

 
1

Summary of Contents for Jewett XR60C

Page 1: ...Installation and Operation Manual Thermo Scientific Jewett Blood Bank Refrigerators ...

Page 2: ... 2008 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries ...

Page 3: ...f electrical characteristics other than those specified on the serial plate will cause permanent damage to the mechanism The serial plate is located inside the refrigerator at the top of the right hand wall 7 All models have been provided with an electric condensate evaporator located on the back of the cabinet Periodic cleaning of the condensate pan will insure proper operation CAUTION PAN AND HE...

Page 4: ...djust any control parameters it is advisable to contact Technical Service before doing so CAUTION Unless a front panel control option has been specified the control panels for BBR blood bank refrigerators are located inside a junction box on the back of the refrigerator To access the control remove the four screws and carefully remove the cover Be sure to avoid touching any component inside the ju...

Page 5: ...CONTROLLING LOADS 3 1 COMPRESSOR The regulation is performed according to the temperature measured by the thermostat probe with a positive differential from the set point if the temperature increases and reaches set point plus differential the compressor is started and then turned off when the temperature reaches the set point value again In case of fault in the thermostat probe the start and stop...

Page 6: ...than 3s the POF message will be displayed 6 7 TO UNLOCK THE KEYBOARD Keep pressed together for more than 3s the 9 9 and 8 8keys till the Pon message will be displayed 6 8 THE CONTINUOUS CYCLE When defrost is not in progress it can be activated by holding the 9 9 key pressed for about 3 seconds The compressor operates in continuous mode for the time set through the CCt parameter The cycle can be te...

Page 7: ...pened after the delay time set through parameter did the door alarm is enabled the display shows the message dA and the regulation restarts The alarm stops as soon as the external digital input is disabled again With the door open the high and low temperature alarms are disabled 8 2 GENERIC ALARM i1F EAL As soon as the digital input is activated the unit will wait for did time delay before signall...

Page 8: ...5 C 0 7 C 1 digit 14 CONNECTIONS 14 1 XR60C 12VAC DV OR 24 VAC DV N C Line 8 3 A250Va 8 3 A Max 16A Comp Def Fan E va p R oo m Hot Key NOTE The compressor relay is 8 3 A or 16 6 A according to the model 24Vac dc supply connect to the terminals 7 and 8 14 2 XR60C 120VAC OR 230 VAC N C Line 8 3 A250Va 8 3 A Max 16A Comp Def Fan E va p R oo m Hot Key NOTE The compressor relay is 8 3 A or 16 6 A accor...

Page 9: ... of slides DRAWER SLIDE MOUNTING INSTRUCTIONS 1 Push into slots in rear standard as shown in Step 1 Make certain that the opposite drawer slides are in the same slots in the opposite standards 2 Push into slots in front standard and press down as shown in Step 2 If necessary tap into place 7 SHELVES OPTIONAL Units for which optional shelves have been ordered come with shelves pre installed at the ...

Page 10: ...e to removal of the dollar bill test strip should be felt Check the perimeter of the door If the latch is loose see latch adjusting information 5 Replace cover plate A Note To accomplish different offsets shims are utilized If replacing hinges make sure to reuse any shims furnished on the cabinet 8 2 Latch and Strike Adjustment 1 Latch is fastened to door 2 For up or down adjustment proper latch e...

Page 11: ...ese RoHS Anweisungfallen finden Sie unter www thermo com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE È marcato col seguente simbolo Thermo Fischer Scientific ha stipulato contratti con una o diverse società di riciclaggio smaltimento in ognuno degli Stati Membri Europei Questo prodo...

Page 12: ...e information can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE SALES DIVISION Phone 1 866 984 3766 1 866 9 THERMO LABORATORY PARTS and SERVICE Phone 1 800 438 4851 TECHNICAL SUPPORT Phone 1 800 438 4851 ...

Page 13: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 306098H01 Rev E ...

Reviews: