Installation and Operation Manual

Thermo Scientific Jewett

®

 

Blood Bank Refrigerators 

Summary of Contents for Jewett XR60C

Page 1: ...Installation and Operation Manual Thermo Scientific Jewett Blood Bank Refrigerators ...

Page 2: ... 2008 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries ...

Page 3: ...f electrical characteristics other than those specified on the serial plate will cause permanent damage to the mechanism The serial plate is located inside the refrigerator at the top of the right hand wall 7 All models have been provided with an electric condensate evaporator located on the back of the cabinet Periodic cleaning of the condensate pan will insure proper operation CAUTION PAN AND HE...

Page 4: ...djust any control parameters it is advisable to contact Technical Service before doing so CAUTION Unless a front panel control option has been specified the control panels for BBR blood bank refrigerators are located inside a junction box on the back of the refrigerator To access the control remove the four screws and carefully remove the cover Be sure to avoid touching any component inside the ju...

Page 5: ...CONTROLLING LOADS 3 1 COMPRESSOR The regulation is performed according to the temperature measured by the thermostat probe with a positive differential from the set point if the temperature increases and reaches set point plus differential the compressor is started and then turned off when the temperature reaches the set point value again In case of fault in the thermostat probe the start and stop...

Page 6: ...than 3s the POF message will be displayed 6 7 TO UNLOCK THE KEYBOARD Keep pressed together for more than 3s the 9 9 and 8 8keys till the Pon message will be displayed 6 8 THE CONTINUOUS CYCLE When defrost is not in progress it can be activated by holding the 9 9 key pressed for about 3 seconds The compressor operates in continuous mode for the time set through the CCt parameter The cycle can be te...

Page 7: ...pened after the delay time set through parameter did the door alarm is enabled the display shows the message dA and the regulation restarts The alarm stops as soon as the external digital input is disabled again With the door open the high and low temperature alarms are disabled 8 2 GENERIC ALARM i1F EAL As soon as the digital input is activated the unit will wait for did time delay before signall...

Page 8: ...5 C 0 7 C 1 digit 14 CONNECTIONS 14 1 XR60C 12VAC DV OR 24 VAC DV N C Line 8 3 A250Va 8 3 A Max 16A Comp Def Fan E va p R oo m Hot Key NOTE The compressor relay is 8 3 A or 16 6 A according to the model 24Vac dc supply connect to the terminals 7 and 8 14 2 XR60C 120VAC OR 230 VAC N C Line 8 3 A250Va 8 3 A Max 16A Comp Def Fan E va p R oo m Hot Key NOTE The compressor relay is 8 3 A or 16 6 A accor...

Page 9: ... of slides DRAWER SLIDE MOUNTING INSTRUCTIONS 1 Push into slots in rear standard as shown in Step 1 Make certain that the opposite drawer slides are in the same slots in the opposite standards 2 Push into slots in front standard and press down as shown in Step 2 If necessary tap into place 7 SHELVES OPTIONAL Units for which optional shelves have been ordered come with shelves pre installed at the ...

Page 10: ...e to removal of the dollar bill test strip should be felt Check the perimeter of the door If the latch is loose see latch adjusting information 5 Replace cover plate A Note To accomplish different offsets shims are utilized If replacing hinges make sure to reuse any shims furnished on the cabinet 8 2 Latch and Strike Adjustment 1 Latch is fastened to door 2 For up or down adjustment proper latch e...

Page 11: ...ese RoHS Anweisungfallen finden Sie unter www thermo com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE È marcato col seguente simbolo Thermo Fischer Scientific ha stipulato contratti con una o diverse società di riciclaggio smaltimento in ognuno degli Stati Membri Europei Questo prodo...

Page 12: ...e information can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE SALES DIVISION Phone 1 866 984 3766 1 866 9 THERMO LABORATORY PARTS and SERVICE Phone 1 800 438 4851 TECHNICAL SUPPORT Phone 1 800 438 4851 ...

Page 13: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 306098H01 Rev E ...