Set up a run
1.
Switch the power switch located at the rear of the unit to the
On
position.
2.
Touch
Set up run
.
3.
In the
Method Type
screen touch either
Constant voltage
,
Constant current
, or
Constant
power
.
Figure 5 Method Type screens
4.
Edit the run parameters (Volts, Time, Amps, and Watts) by touching the parameter on the screen to
bring up the
Run parameter edit
screen.
Default maximum values are 500 V, 3000 mA, and 600 W.
Note:
When using constant voltage or constant power, it is recommended that the maximum
default settings not be changed to avoid unwanted power limitations during the run.
5.
Enter the parameter value required, then touch
Enter
to advance with the newly changed run
parameter displayed.
6.
(Optional)
Add or delete extra steps by touching
Actions
.
Chapter 2
Methods
Set up a run
2
12
PowerEase
™
Touch 600W Power Supply User Guide