iCAP 6000 Series 
Pre-Installation Manual 

Version 4.2 
  
Part number 8499 400 30001 
 

Summary of Contents for iCAP 6000 Series

Page 1: ...served Registration No 441506 SOLAAR House 19 Mercers Row Cambridge CB5 8BZ United Kingdom Telephone 44 0 1223 347400 Fax 44 0 1223 347402 http www thermofisher com iCAP 6000 Series Pre Installation Manual Version 4 2 Part number 8499 400 30001 ...

Page 2: ...Thermo Fisher Scientific iCAP 6000 Series Pre Installation Manual v4 2 ii This page is intentionally blank ...

Page 3: ...equirements 4 1 Introduction 4 1 Location 4 1 Environmental Requirements 4 2 Electrical 4 2 Spectrometer Size 4 4 Gas Requirements 4 5 Water Temperature Control 4 5 Waste Storage 4 6 Spectrometer Fume Extraction 4 6 Sample Fume Extraction 4 8 Communications Interface 5 9 On the Spectrometer 5 9 On the Data Station 5 9 Pre Installation Checklist 6 11 Special Agreements 7 1 Chapter 1 Chapter 2 Chapt...

Page 4: ...iv iCAP 6000 Series Pre Installation Manual v4 2 Thermo Fisher Corporation This page is intentionally blank ...

Page 5: ...lation are also detailed WARNING The installation of all services must comply with the appropriate rules and regulations required by the local authorities responsible for those services in the workplace at the installation site An installation engineer is not responsible for the fitting or compliance of the facilities or services The choice of an operating site for the instrument will be influence...

Page 6: ...RS232 on Data Station Cetac ASX 1400 Autosampler Stirring Cetac U 6000AT Ultrasonic Nebuliser Elemental Scientific Inc SC SC4 Autosampler Elemental Scientific Inc SC4 Fast Autosampler Argon Humidifier SSEA Solid sampling device requires dedicated RS232 on Data Station and iCAP 6000 accessory board Laser Solid sampling device requires dedicated RS232 on Data Station and iCAP 6000 accessory board Hi...

Page 7: ... the customer receiving area to the installation site must be checked for suitability The largest pallet size should be used to calculate clearances WARNING The boxed Spectrometer should be moved using a mechanical lift Five people should be used if it has to be moved by hand taking appropriate precautions Care should be taken in doorways corridors or when lowering to avoid trapping fingers The in...

Page 8: ...damage The nature of the damage should be noted on the delivery notice and signed by the carrier s representative Within the time stated in the Thermo Fisher terms and conditions an inspection should be made for concealed damage The local Thermo Fisher organisation should be advised of any damage in writing and on receipt of specific instructions the customer should return the equipment complete a...

Page 9: ...ration Particular consideration should be taken to avoid direct sunlight proximity to heat sources draughts and vibration Do not locate the system where sudden changes in temperature can occur for example near a door or window care should be taken with the location of items such as air conditioning vents and heating vents The instruments are designed for use on a normal laboratory bench Ideally th...

Page 10: ...extraction is turned off sufficient air must be supplied to prevent the concentration of Argon reaching a harmful level The spectrometer will require an electrical supply at 200 to 240 VAC 4 KVA 50 60 Hz The spectrometer will be supplied with a 2 5m mains cord For customers in the USA and Canada the mains cord supplied will terminate in a NEMA L6 20P Twist Lock plug Coding is black X ac high white...

Page 11: ...d mains sockets will be required for the Data Station PC and for an autosampler chiller printer or any other additional accessory required Each electrical outlet must have an effective earth ground connection This protection must not be negated by the use of an extension cable without a protective earth conductor ...

Page 12: ... v3 1 Thermo Fisher Corporation Width Depth Height Weight 832 mm 744 mm 589 mm 85 5 Kg 32 7 in 29 3 in 23 2 in 188 lbs Table 4 2 Spectrometer Dimensions and Weight 831 2mm Figure 4 2 iCAP 6000 Series Front View 510mm Figure 4 3 iCAP 6000 Series Top View Spectrometer Size ...

Page 13: ... purge and plastic for the plasma gas Connection to the laboratory gas supply should be within 1 metre of the instrument A regulator should be close to the connection Care should be taken to ensure that laboratory gas lines do not contaminate the gases used for the iCAP If liquid argon is not used it is recommended that banks of cylinders be used with switch over valves so that the gas supplies ca...

Page 14: ...e provided for the disposal of any waste which should be disposed of following local procedures and regulations This instrumentation is designed for operation in clean air conditions The laboratory must be free of all contaminants that could have a degrading effect on the instrument components Dust acid and organic vapours must be excluded from the work area The warranty will be void if the equipm...

Page 15: ...t be adjustable to 50 of the maximum to allow optimization during installation This can be achieved with a butterfly valve or an adjustable gate as shown in figure 4 3 below The extraction system must not be affected by external weather conditions or other uses the system may be used for Figure 4 4 iCAP 6000 Extraction System PLEASE NOTE EXTRACTION REQUIREMENTS ARE NOT THE SAME AS FOR PREVIOUS THE...

Page 16: ...stallation Manual v3 1 Thermo Fisher Corporation Additional separate extraction should be considered if significant numbers of volatile or acidic samples are left in the proximity of the instrumentation even those in an autosampler Sample Fume Extraction ...

Page 17: ...r each accessory The Data Station requirements will normally be met by a PC meeting the minimum specification Windows XP Professional Service pack 2 20Gb Hard drive 512 Mb RAM Dedicated Ethernet card for spectrometer communication Some accessories require a dedicated RS232 port A separate Ethernet card will be required for networking A printer port maybe required A suitable PC is available from yo...

Page 18: ...Communications Interface 5 10 iCAP 6000 Series Pre Installation Manual v3 1 Thermo Fisher Corporation This page is intentionally blank ...

Page 19: ...d Completed Location Table dimensions for the spectrometer 900 x 600 mm Table dimension for the auto sampler right side of the machine 600 x 550 mm Extraction Tube diameter for extraction 125 mm Extraction capability 10 m s Duo 5 m s Radial Adjustable valve in the extraction tube Yes Electrical supply Plug 200 240 VAC 32A single phase 4000W 50Hz Yes Standard mains plug 200 240 VAC for each accesso...

Page 20: ... for the rinse supply station and waste 2 needed Chiller if not bought with the instrument Maximum distance from the spectrometer 3 m Pipe supply O D dimension 12 mm Cooling capability 900 W Flow rate 5 l min Chiller inhibitor Yes PC and printer if not bought with the machine 512 Mb RAM 40 Gb HDD 2 Ethernet ports 1 serial port USB ports monitor ink jet printer Yes Windows XP Professional Service p...

Page 21: ...her Scientific iCAP 6000 Series Pre Installation Manual v 4 2 7 1 Chapter 7 Special Agreements This section should detail any special agreements which have been arranged with your local Thermo Fisher organisation ...