Before Use
Leveling the Centrifuge
Thermo Scientific
Heraeus Pico 17 / 21
2-3
•
The location must be free of grease and dust.
•
The centrifuge should not be exposed to heat and strong sunlight.
•
The set-up location must be well-ventilated at all times.
Leveling the Centrifuge
The horizontal leveling of the centrifuge must be checked every time after moving it to a different
location.
The supporting structure must be suitable for horizontal setup of the centrifuge.
Mains Connection
Figure 2-2.
Mains Connection
1.
Turn off the power supply switch on the back (press
“
0
”
).
2.
Plug the centrifuge into grounded electrical sockets only.
3. Check whether the cable complies with the safety standards of your country.
WARNING
UV rays reduce the stability of plastics.
Do not subject the centrifuge, rotors and plastic accessories to direct sunlight.
CAUTION
If the centrifuge is not level, imbalances can occur and the centrifuge can be
damaged.
Do not place anything under the centrifuge to level.
Note
The centrifuge must be connected directly to the socket. Extension cords and multiple
sockets are not allowed.
Main Power Switch
Mains plug