1
Thermo Scientific
1
Operating environment
Note
The electromagnetic environment should be evaluated prior to operation of the
analyzer. Do not use this analyzer in close proximity to sources of strong electromagnetic
radiation, for example, unshielded intentional RF-sources. They may interfere with the proper
operation.
Note
Installing any 3
rd
party software to the workstation PC is not supported by Thermo
Fisher Scientific. The 3
rd
party software (e.g. antivirus software) can interfere with the analyzer
performance and, for instance, generate error messages like “internal timing error” or “master
used too much time”. Thermo Fisher Scientific is not responsible for possible errors due to 3
rd
party software installed by customer.
Note
It is not recommended to connect the workstation to the Internet due to the risk of
viruses, unless specifically requested by Thermo Fisher Scientific in order to join a remote
support session. To use remote connection, refer to the instructions in the Operation Manual.
Dimensions
Table 2.
Dimensions and weight of the analyzer
Description
cm / kg
in / lb
Width
94 cm
37.0 in
Height
62 cm / 130 cm
24.4 in / 51.2 in
Depth
70 cm
27.6 in
Weight
110 kg
242.5 lb
Note
Dimensions are given without workstation and display.
Power supply
Table 3.
Power supply to the analyzer
Description
Value
Voltage
100 - 240 VAC ± 10%
Frequency
50 - 60 Hz ± 5%
Summary of Contents for Gallery Plus
Page 1: ......
Page 6: ...vi Thermo Scientific Figure 2 Safety labeling inside analyzer ...
Page 12: ... this page was intentionally left blank ...
Page 18: ... this page was intentionally left blank ...
Page 26: ... this page was intentionally left blank ...
Page 44: ...4 Setting up Setting up 26 Thermo Scientific 7 Connect the workstation s power cable ...
Page 47: ...A Thermo Scientific 29 Removing transport casing ...
Page 48: ... this page was intentionally left blank ...
Page 52: ... this page was intentionally left blank ...