 Installation  |   7

Door ramp alignment feature

Opening the Door

1. Remove the padlock if installed.

2. Grasp the latch handle and pull it toward yourself until the 

latch disengages.

3. Keep pulling by the latch handle to open the main door.

Closing the Door

Note: 

The latch does not self-engage automatically when you 

close the door. You must rotate the latch into the open 
position first.

1. Grasp the latch handle and pull it toward yourself, 

rotating the latch into the open position.

2. Move the freezer door into the closed position and gently 

push the handle away from you, making sure that the 
latch engages fully with the cabinet strike.

3. Keep applying gentle pressure to the latch handle until 

the latch is securely in closed position.

4. Insert the key and rotate counterclockwise to lock.

5. Replace the padlock as required.

Pressure Equalization Port

When an upright ultra-low temperature freezer door is 
opened, room temperature air rushes into the storage 
compartment. When the door is closed, the fixed volume of 
air is cooled rapidly. Pressure drops below atmospheric 
pressure, resulting in a substantial vacuum. Re-entry into the 
cabinet is impossible until internal pressures are returned to 
atmospheric pressure. Without a pressure equalization 
mechanism, it can take, in extreme cases, several hours 
before the door can easily be reopened.

All upright models feature a port that provides vacuum relief 
after door openings. The pressure equalization port is located 
in the door behind the eye-level panel on the front of the 
freezer. Although the port is heated and designed to self-
defrost, excessive frost accumulation on the inner door could 
eventually restrict air flow. Therefore, you should periodically 
inspect the inner door and brush away any loose frost using a 
stiff nylon brush.

Installing the Remote 
Alarm Connector

The remote alarm contacts are located on the back of the 
freezer above and to the left of the power switch. After 
installing the wiring from the remote alarm to the connector, 
install the connector to the freezer micro-board.

The pin configuration is shown in Figure 2 below.

Figure 2. Remote Alarm Pin Configuration

For systems that alarm in closed state, connect to pins 5 & 6.
For systems that alarm in open state, connect to pins 6 & 7.

The contacts will trip in the event of a power outage, high 
temperature alarm, low temperature alarm or door ajar alarm.

Intended Use

The –86°C freezers (refer to Models for the specific model 
series) described in this manual are intended for the storage 
of Frozen Red Blood Cells in 40% glycerol at -65°C for up to 
10 years; Frozen Rejuvenated Red Blood Cells at -65°C for 
up to 10 years; Fresh Frozen Plasma at -65°C for up to 10 
years; and Frozen or Cryopreserved Tissue for long term 
storage: Musculoskeletal Tissue at -40°C or colder, Skin at -
40°C or colder and Bone at -40°C or colder.

These products should only be used by an authorized and 
adequately trained personnel. It is considered a medical 
device and has therefore been listed with medical device 
regulatory body FDA and is considered a Class II medical 
device by the FDA. This product is listed under product code 
KSE, regulation number 864.9700 and is considered 510(K) 
exempt. This unit is not intended for use in classified 
hazardous locations, nor to be used for the storage of 
flammable inventory.

CAUTION: 

When moving the freezer, always 

grasp cabinet surfaces; never pull the freezer by 
the latch handle.

Summary of Contents for Forma FDE Series

Page 1: ...Ultra Low Temperature Freezers RevcoTM RDE Series FormaTM FDE Series HERAfreezeTM HDE Series and Thermo ScientificTM TDE Series Installation and Operation 329712H02 Revision A March 2019...

Page 2: ...med by qualified service personnel Material in this manual is for informational purposes only The contents and the product it describes are subject to change without notice Thermo Fisher Scientific ma...

Page 3: ...attery Door Opening Closing 4 Operating Standards 5 Electrical Specifications 5 Installation 6 Location 6 Leveling 6 Backup System Optional 6 Super Insulated Cabinet Construction 6 Door Operation 6 Pr...

Page 4: ...on 14 Changing Chart Paper 14 Calibration Adjustment 14 Maintenance 15 Cleaning the Condenser 15 Cleaning the Condenser Filter 15 Gasket Maintenance 15 Defrosting the Freezer 15 Battery Maintenance 15...

Page 5: ...odels 1 Models Table 1 Applicable Models Brand Model Size xxx Voltage Forma FDExxx86L 300 400 500 600 A D V Thermo Scientific TDExxx86L 300 400 500 600 A D V HERAfreeze HDExxx86L 300 400 500 600 A D V...

Page 6: ...ntamination procedures use chemically resistant gloves Use insulated gloves for handling samples and when using liquid nitrogen Before installing using or maintaining this product please be sure to re...

Page 7: ...en submitting a claim for shipping damage request that the carrier inspect the shipping container and equipment The packaging can be stored and re used Packing List Inside the freezer cabinet is a bag...

Page 8: ...helf allow the freezer to recover to the desired set point before loading the next shelf Repeat this process until the freezer is fully loaded Battery Door Opening Closing To open the grille door pull...

Page 9: ...he nominal voltage Electrical Specifications The last character in the model number listed on the data plate identifies the electrical specifications for your unit Specific unit current rating is list...

Page 10: ...odels are equipped with an advanced assembly specifically designed for ultra low temperature freezers Features include One hand operation A front accessible lock Hasps for a standard padlock to provid...

Page 11: ...excessive frost accumulation on the inner door could eventually restrict air flow Therefore you should periodically inspect the inner door and brush away any loose frost using a stiff nylon brush Ins...

Page 12: ...273 274 STANDARDS FOR TISSUE BANKING 14th Ed Section E E3 300 E3 310 Table E3 331 Note for clarity The Bone is the Osteoarticular Graft a weight bearing allograft with intact articular surfaces consi...

Page 13: ...larm value The range is 99 C to within 5 C of setpoint Note A setpoint change may automatically change the warm cold alarm setpoints as well to maintain a minimum 5 C separation from the control setpo...

Page 14: ...will automatically return to the temperature display Warm Alarm Test Press the plus and checkmark buttons at the same time to initiate the warm alarm test During the warm alarm test the actual cabine...

Page 15: ...are showing Press the checkmark while YES is showing With a flashing YES you are required to confirm by pressing the checkmark button a second time Power down will then be complete If the checkmark b...

Page 16: ...equipped with siphon tubes for withdrawing liquid from the bottom of the cylinder CO2 cylinders must be kept at room temperature to function properly LN2 bottles are functional at any reasonable temp...

Page 17: ...ay shorten the life of the hose Start Up When the unit is started it will recognize if a backup system is installed 1 Follow the instructions in Backup System Optional to set the backup system type an...

Page 18: ...econd The pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the reference mark a small groove on the lef...

Page 19: ...et 6 If there is freezer odor wash the interior with a solution of baking soda and warm water 7 Clean the exterior with any common household cleaner 8 Close the doors restart the freezer and reload Re...

Page 20: ...und the gasket inner doors and breakers as necessary Gasket Periodically check the gaskets around the door for punctures or tears Periodically clean the ice build up around the gasket Filter Clean the...

Page 21: ...igh relative humidity Check if the location meets requirements Maximum relative humidity 60 for temperatures within 15 C to 32 C 59 F to 90 F Excess frost build up in chamber Defrost the unit Refer De...

Page 22: ...o puncture the rubber gasket Excessive and prolonged door openings Avoid opening door for a prolonged time Gasket damage Check for punctures or tears on gasket If replacement is necessary call service...

Page 23: ...are running more often than before Ambient conditions Unit performance is directly impacted by these causes mentioned Try maintaining ambient conditions reducing load reducing door openings Once temp...

Page 24: ...01 Firmware Build Incompatible E02 Control Probe Failure E03 Heat Exchanger Probe Failure E04 Power Failure E05 Failure to Reach Setpoint E06 BUS Battery Failure E07 System Battery Failure E08 Lost Co...

Page 25: ...luded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original fi...

Page 26: ...able items i e glass filters pilot lights light bulbs and door gaskets are excluded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the wa...

Page 27: ...rature variation will increase Contact customer service Display will intermittently show E03 Ambient Probe Failure Wrench Yes Ambient Probe TC has malfunctioned This doesn t affect the performance of...

Page 28: ...s incompatibility that can result in modules to be non coherent Display will intermittently show E01 System Battery Failure Alarm Wrench Yes System Battery Disconnected or Failed Display will intermit...

Page 29: ...thermofisher com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2012 19 EU in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE marcato col...

Page 30: ...contact information for your area Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States Countries Sales North America 1 866 984 3766 India toll free 1800 22 8374 India 91 22 671...