12  |   Backup System (Optional)

Thermo Fisher Scientific ULT Freezers

Backup System (Optional)

For all ultra-low temperature cold storage products, we 
recommend the use of a backup system (BUS) for the 
security of your samples.

When you purchase a built-in CO

2

 or LN

2

 optional backup 

system for the freezer, backup control is integrated into the 
main user interface.

Note: 

For stand-alone backup systems, refer to installation 

instructions provided with the backup system kit.

CO

2

 and LN

2

 Precautions

The following are precautions for using liquid CO

2

 and LN

2

 

backup systems.

Installation

Field installed systems are supplied with complete installation 
and operating instructions. If your system is factory installed, 
the freezer is shipped with a coiled length of hose to connect 
the freezer to the bottles:

1/4” Flexible Hose with fittings for connection to the CO

2

supply.

1/2” Flexible Hose with fittings for connection to the LN

2

supply.

To install,

1. Straighten the coiled hose.

2. Connect one end to the labeled connection on the

freezer.

Tighten the nut two flats past finger tight, approximately
120 degrees.

Note: 

For CO

2

, remove the threaded fitting from the nut on 

the end of the copper tubing to access nut for connection to 
the freezer. Discard the threaded fitting.

3. Attach the other end to the supply bottle or building

supply fitting.

For  CO

2

:

Remove Nipple from adapter (NPT Connection).
Remove cable tie to release alternative nut and
washer. Ensure the correct nut fitting is supplied over
the nipple (US or European).

Note: 

Always purchase the cylinders which are 

equipped with siphon tubes for withdrawing liquid 
from the bottom of the cylinder. CO

2

 cylinders 

must be kept at room temperature to function 
properly. LN

2

 bottles are functional at any 

reasonable temperature.

WARNING: 

If a CO

2

 or LN

2

 cylinder falls and a 

valve is knocked off, the cylinder becomes a 
deadly and completely unguided missile. 
Transport the cylinders in a hand-truck or cart 
with secure chain ties for the cylinder. After 
cylinders are connected to the equipment, 
securely attach them with chains to a solid 
stationary object such as a building column.

WARNING: 

CO

2

 and LN

2

 liquids are non-

poisonous but are very cold and will burn 
unprotected skin. Always wear protective eye 
wear and clothing when changing cylinders or 
working on the piping systems attached to an 
active source of liquid refrigerant.

WARNING: 

The gases produced by evaporation 

of CO

2

 or LN

2

 are non-poisonous but displace 

the oxygen in a confined space and can cause 
asphyxiation. Do not store the cylinders in 
subsurface or enclosed areas.

CAUTION: 

When closing the cylinder valve, 

make sure that the injection solenoid is energized 
to allow all the liquid to bleed off instead of being 
trapped in the supply hose. Failure to do this 
results in activation of the pressure relief device, 
which could damage the freezer and requires 
replacing if it is activated.

CAUTION: 

For models ordered with factory 

installed built-in backup systems, the flow of 
liquid CO

2

 or LN

2

 will be discontinued if the door 

is opened during operation of the backup 
system. For units operated with free-standing, 
field installed type backup system, the flow of 
liquid CO

2

 or LN

2

 will be discontinued upon door 

opening only if the switch provided with the free-
standing package is installed on the freezer.

Summary of Contents for Forma FDE Series

Page 1: ...Ultra Low Temperature Freezers RevcoTM RDE Series FormaTM FDE Series HERAfreezeTM HDE Series and Thermo ScientificTM TDE Series Installation and Operation 329712H02 Revision A March 2019...

Page 2: ...med by qualified service personnel Material in this manual is for informational purposes only The contents and the product it describes are subject to change without notice Thermo Fisher Scientific ma...

Page 3: ...attery Door Opening Closing 4 Operating Standards 5 Electrical Specifications 5 Installation 6 Location 6 Leveling 6 Backup System Optional 6 Super Insulated Cabinet Construction 6 Door Operation 6 Pr...

Page 4: ...on 14 Changing Chart Paper 14 Calibration Adjustment 14 Maintenance 15 Cleaning the Condenser 15 Cleaning the Condenser Filter 15 Gasket Maintenance 15 Defrosting the Freezer 15 Battery Maintenance 15...

Page 5: ...odels 1 Models Table 1 Applicable Models Brand Model Size xxx Voltage Forma FDExxx86L 300 400 500 600 A D V Thermo Scientific TDExxx86L 300 400 500 600 A D V HERAfreeze HDExxx86L 300 400 500 600 A D V...

Page 6: ...ntamination procedures use chemically resistant gloves Use insulated gloves for handling samples and when using liquid nitrogen Before installing using or maintaining this product please be sure to re...

Page 7: ...en submitting a claim for shipping damage request that the carrier inspect the shipping container and equipment The packaging can be stored and re used Packing List Inside the freezer cabinet is a bag...

Page 8: ...helf allow the freezer to recover to the desired set point before loading the next shelf Repeat this process until the freezer is fully loaded Battery Door Opening Closing To open the grille door pull...

Page 9: ...he nominal voltage Electrical Specifications The last character in the model number listed on the data plate identifies the electrical specifications for your unit Specific unit current rating is list...

Page 10: ...odels are equipped with an advanced assembly specifically designed for ultra low temperature freezers Features include One hand operation A front accessible lock Hasps for a standard padlock to provid...

Page 11: ...excessive frost accumulation on the inner door could eventually restrict air flow Therefore you should periodically inspect the inner door and brush away any loose frost using a stiff nylon brush Ins...

Page 12: ...273 274 STANDARDS FOR TISSUE BANKING 14th Ed Section E E3 300 E3 310 Table E3 331 Note for clarity The Bone is the Osteoarticular Graft a weight bearing allograft with intact articular surfaces consi...

Page 13: ...larm value The range is 99 C to within 5 C of setpoint Note A setpoint change may automatically change the warm cold alarm setpoints as well to maintain a minimum 5 C separation from the control setpo...

Page 14: ...will automatically return to the temperature display Warm Alarm Test Press the plus and checkmark buttons at the same time to initiate the warm alarm test During the warm alarm test the actual cabine...

Page 15: ...are showing Press the checkmark while YES is showing With a flashing YES you are required to confirm by pressing the checkmark button a second time Power down will then be complete If the checkmark b...

Page 16: ...equipped with siphon tubes for withdrawing liquid from the bottom of the cylinder CO2 cylinders must be kept at room temperature to function properly LN2 bottles are functional at any reasonable temp...

Page 17: ...ay shorten the life of the hose Start Up When the unit is started it will recognize if a backup system is installed 1 Follow the instructions in Backup System Optional to set the backup system type an...

Page 18: ...econd The pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install new chart paper Carefully align the day and time with the reference mark a small groove on the lef...

Page 19: ...et 6 If there is freezer odor wash the interior with a solution of baking soda and warm water 7 Clean the exterior with any common household cleaner 8 Close the doors restart the freezer and reload Re...

Page 20: ...und the gasket inner doors and breakers as necessary Gasket Periodically check the gaskets around the door for punctures or tears Periodically clean the ice build up around the gasket Filter Clean the...

Page 21: ...igh relative humidity Check if the location meets requirements Maximum relative humidity 60 for temperatures within 15 C to 32 C 59 F to 90 F Excess frost build up in chamber Defrost the unit Refer De...

Page 22: ...o puncture the rubber gasket Excessive and prolonged door openings Avoid opening door for a prolonged time Gasket damage Check for punctures or tears on gasket If replacement is necessary call service...

Page 23: ...are running more often than before Ambient conditions Unit performance is directly impacted by these causes mentioned Try maintaining ambient conditions reducing load reducing door openings Once temp...

Page 24: ...01 Firmware Build Incompatible E02 Control Probe Failure E03 Heat Exchanger Probe Failure E04 Power Failure E05 Failure to Reach Setpoint E06 BUS Battery Failure E07 System Battery Failure E08 Lost Co...

Page 25: ...luded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original fi...

Page 26: ...able items i e glass filters pilot lights light bulbs and door gaskets are excluded from this warranty Replacement or repair of component parts or equipment under this warranty shall not extend the wa...

Page 27: ...rature variation will increase Contact customer service Display will intermittently show E03 Ambient Probe Failure Wrench Yes Ambient Probe TC has malfunctioned This doesn t affect the performance of...

Page 28: ...s incompatibility that can result in modules to be non coherent Display will intermittently show E01 System Battery Failure Alarm Wrench Yes System Battery Disconnected or Failed Display will intermit...

Page 29: ...thermofisher com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2012 19 EU in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE marcato col...

Page 30: ...contact information for your area Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States Countries Sales North America 1 866 984 3766 India toll free 1800 22 8374 India 91 22 671...

Reviews: