Chapter 2
The FOCUS GC User Interface
Configuration Menu
54
Carrier Configuration Menu
Commonly used gases are Helium, Hydrogen and Nitrogen. Hydrogen offers the
best chromatography properties (optimum resolution at the highest flow rates).
However, because Hydrogen is flammable, Helium is often used as an alternative
for carrier gas.
You specify the type of carrier gas in the
CONFIGURE CARRIER
menu by using
the
INC
and
DEC
key.
WARNING! Hydrogen items is disabled unless your instrument has a hydrogen sensor. See
CONFIGURE CARRIER
Gas Type Helium
Table 2-5. Configure Carrier Menu
Parameter
Range
Comments
CONFIGURE CARRIER
This line is the title bar.
Gas type
Helium
Nitrogen
Hydrogen
Argon
To select the required carrier gas. The default carrier
gas is Helium (He).
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...