9
8 Segment & 4x16 Segment Programmable Models w/OTP
Controller Parameters
34
Large Tabletop Muffle
Thermo Scientific
Controller Parameters
Home Display
°C
: measured temperature in Celsius.
Temperature units cannot be changed without entering the configuration. Contact Customer Service if
a different temperature unit is required.
OP
: % output power demand; displayed in lower display (cannot be changed).
C.id
: Controller identification number.
PrG
: Program number (displayed when a program is running; 4x16 programmable models only).
IdHi
: Deviation High Alarm.
tunE
: One-shot autotune enable.
run LiSt
(Program Run List)
PrG
: Currently running program (only used on 4x16 programmable models).
StAt
: Displays the program status [OFF, run (running active program), hoLd (program on hold),
HbAc (waiting for process to catch up), End (program completed)] in the lower display. The controller
will default to “OFF.”
FASt
: Fast run through program (no/YES). The controller will default to “no.”
SEG.d
: Flash active segment type in the lower display of the home display (no/YES). The controller
will default to “no.”
ProG LiSt (Program Edit List)
PrG.n
: Press the UP or DOWN ARROW to select the program number (program number will be
displayed in lower display on 4x16 programmable models only).
Hb
: Press the UP or DOWN ARROW to select the holdback type [OFF (disables holdback), Lo
(deviation low holdback), Hi (deviation high holdback) or bAnd (deviation band holdback)] for the
entire program. The controller will default to “OFF.”
Hb.U
: Press the UP or DOWN ARROW to select the holdback value (in display units).
rmP.U
: Press the UP or DOWN ARROW to toggle between ramp units (Sec, min or Hour).
Controller will default to “Sec.”
dwL.U
: Press the UP or DOWN ARROW to toggle between dwell units (SEc, min or Hour).
Controller will default to “SEc.”
Cyc.n
: Press the UP or DOWN ARROW to set the number of program cycles (1 to 999 or cont). The
controller will default to “cont.”
SEG.n
: Press the UP or DOWN ARROW to select the segment number (1-8 in 8 segment models,
1-16 in 4x16 models).
Summary of Contents for F30400
Page 6: ...Contents 4 Large Tabletop Muffle Thermo Scientific...
Page 10: ...1 Safety Information Warning 8 Large Tabletop Muffle Thermo Scientific...
Page 14: ...2 Introduction Principles of Operation 12 Large Tabletop Muffle Thermo Scientific...
Page 18: ...4 Unpacking 16 Large Tabletop Muffle Thermo Scientific...
Page 28: ...7 Single Setpoint Models w OTP Tuning 26 Large Tabletop Muffle Thermo Scientific...
Page 50: ...11 Furnace Loading 48 Large Tabletop Muffle Thermo Scientific...
Page 62: ...15 Replacement Parts List 60 Large Tabletop Muffle Thermo Scientific...
Page 64: ...16 Ordering Procedures 62 Large Tabletop Muffle Thermo Scientific...