6.
Open the
Oxygen sensor
4
Calibrate
controls
1
from inside the
Incubator
window.
Common Callouts and Arrows
1. Copy-paste a callout or arrow to use in this SVG.
Note: If you need more advanced callouts or arrows
use the TechComm_Inkscape_Callout&Arrow_Libary.
3. Delete this text, this rectangle, and unused
callouts, arrows, or other SVG elements
before adding this SVG to the repository.
2. Edit number and/or line-length, as needed.
1
1
1
1
1
2
3
4
1
Calibrate controls
2
Oxygen %
3
Purge gas selection
4
Begin Calibration button
7.
Set the
Oxygen %
2
to the oxygen content of the air used by the system.
Note:
The default value for
Oxygen %
is 20.95% for normal compressed air.
8.
Select the purge gas to use:
Nitrogen
or
CO
2
3
.
9.
Press
Begin Calibration
4
to start the calibration routine.
The calibration process takes approximately three minutes to complete. When complete, the system displays the
“Calibration was
successful."
message next to the
Begin Calibration
button.
If there was an error during calibration, the system displays the
"Calibration error, please check if gas is connected."
message.
10.
In case of a calibration error, verify that the gas is connected, the gas valves and regulators are on, and the correct gas was selected as the
purge gas, then repeat the calibration routine.
Note:
If the calibration error persists, contact Technical Support.
11.
Press
Done
when finished.
4
EVOS
™
Onstage Incubator O
2
Sensor Replacement Kit