background image

EVOS

 Onstage Incubator O

2

 Sensor Replacement Kit

Catalog Number  

AMEP4947

Doc. Part No.  

 710399(01)     

Pub. No.  

 MAN0025857   

Rev.  

 A.0

Note: 

For safety and biohazard guidelines, see the "Safety" appendix in the 

EVOS

 M7000 Imaging System User Guide

 (Pub. No. MAN0018326)

or the 

EVOS

 M5000 Imaging System User Guide

 (Pub. No. MAN0017563). Read the Safety Data Sheets (SDSs) and follow the handling

instructions. Wear appropriate protective eyewear, clothing, and gloves.

Kit contents

• O

2

 sensor

• O

ring

• 2

mm Allen wrench

• Instructions

Replace the O

2

 sensor

1.

Power off the EVOS

 Onstage Incubator, then unplug the power cord, the USB cable, and the 6-pin sensor data cable from the control unit of

the incubator.

2.

Turn off the regulators on the gas tanks, then detach the gas lines from the control unit.

3.

Use the 2

mm Allen key to remove the 3 button head screws, then slide out the side panel.

 

 

Common Callouts and Arrows

1.  Copy-paste a callout or arrow to use in this SVG.
 
     Note:  If you need more advanced callouts or arrows
     use the TechComm_Inkscape_Callout&Arrow_Libary.

3.  Delete this text, this rectangle, and unused
     callouts, arrows, or other SVG elements
     before adding this SVG to the repository.

2.  Edit number and/or line-length, as needed.

1

1

1

1

 

4.

Unplug the O

2

 sensor's connector (

1

: plugged, 

2

: unplugged).

 

Common Callouts and Arrows

1.  Copy-paste a callout or arrow to use in this SVG.
 
     Note:  If you need more advanced callouts or arrows
     use the TechComm_Inkscape_Callout&Arrow_Libary.

3.  Delete this text, this rectangle, and unused
     callouts, arrows, or other SVG elements
     before adding this SVG to the repository.

2.  Edit number and/or line-length, as needed.

1

1

1

1

 

1

2

 

USER GUIDE

For Research Use Only. Not for use in diagnostic procedures.

2 November 2021

Summary of Contents for EVOS AMEP4947

Page 1: ...rotective eyewear clothing and gloves Kit contents O2 sensor O ring 2 mm Allen wrench Instructions Replace the O2 sensor 1 Power off the EVOS Onstage Incubator then unplug the power cord the USB cable and the 6 pin sensor data cable from the control unit of the incubator 2 Turn off the regulators on the gas tanks then detach the gas lines from the control unit 3 Use the 2 mm Allen key to remove th...

Page 2: ... not apply excessive pressure to prevent damage to the sensor 6 Verify that the O ring 1 is installed on the new O2 sensor If not install the O ring as shown in the image 1 7 Screw in the new O2 sensor until the O ring is compressed between the clear manifold block and the O2 sensor 2 EVOS Onstage Incubator O2 Sensor Replacement Kit ...

Page 3: ...ensor data cable from the environmental chamber into the appropriate input jack on the control unit 3 Re attach the gas lines Air and Nitrogen or CO2 via the PTC push to click connectors to the control unit then turn on the valves on the gas tanks and regulators 4 First turn on the EVOS Onstage Incubator then power on the EVOS M7000 Imaging System 5 Start the EVOS M7000 Software navigate to the Se...

Page 4: ...ation process takes approximately three minutes to complete When complete the system displays the Calibration was successful message next to the Begin Calibration button If there was an error during calibration the system displays the Calibration error please check if gas is connected message 10 In case of a calibration error verify that the gas is connected the gas valves and regulators are on an...

Page 5: ...turn on the EVOS Onstage Incubator then power on the EVOS M5000 Imaging System 5 Start the EVOS M5000 Software click the Settings button 1 in the upper right of the screen then click to open the Incubator menu 2 1 2 3 1 Settings button 2 Incubator menu 3 Oxygen Sensor Calibration button 6 Click Oxygen Sensor Calibration 3 in the Incubator menu to open the Incubator Oxygen Sensor Calibration window...

Page 6: ... any questions please contact Life Technologies at www thermofisher com support thermofisher com support thermofisher com askaquestion Life Technologies Corporation 22025 20th Ave SE Ste 100 Bothell Washington 98021 USA For descriptions of symbols on product labels or product documents go to thermofisher com symbols definition The information in this guide is subject to change without notice DISCL...

Reviews: