EVOS
™
Onstage Incubator O
2
Sensor Replacement Kit
Catalog Number
AMEP4947
Doc. Part No.
710399(01)
Pub. No.
MAN0025857
Rev.
A.0
Note:
For safety and biohazard guidelines, see the "Safety" appendix in the
EVOS
™
M7000 Imaging System User Guide
(Pub. No. MAN0018326)
or the
EVOS
™
M5000 Imaging System User Guide
(Pub. No. MAN0017563). Read the Safety Data Sheets (SDSs) and follow the handling
instructions. Wear appropriate protective eyewear, clothing, and gloves.
Kit contents
• O
2
sensor
• O
‑
ring
• 2
‑
mm Allen wrench
• Instructions
Replace the O
2
sensor
1.
Power off the EVOS
™
Onstage Incubator, then unplug the power cord, the USB cable, and the 6-pin sensor data cable from the control unit of
the incubator.
2.
Turn off the regulators on the gas tanks, then detach the gas lines from the control unit.
3.
Use the 2
‑
mm Allen key to remove the 3 button head screws, then slide out the side panel.
Common Callouts and Arrows
1. Copy-paste a callout or arrow to use in this SVG.
Note: If you need more advanced callouts or arrows
use the TechComm_Inkscape_Callout&Arrow_Libary.
3. Delete this text, this rectangle, and unused
callouts, arrows, or other SVG elements
before adding this SVG to the repository.
2. Edit number and/or line-length, as needed.
1
1
1
1
4.
Unplug the O
2
sensor's connector (
1
: plugged,
2
: unplugged).
Common Callouts and Arrows
1. Copy-paste a callout or arrow to use in this SVG.
Note: If you need more advanced callouts or arrows
use the TechComm_Inkscape_Callout&Arrow_Libary.
3. Delete this text, this rectangle, and unused
callouts, arrows, or other SVG elements
before adding this SVG to the repository.
2. Edit number and/or line-length, as needed.
1
1
1
1
1
2
USER GUIDE
For Research Use Only. Not for use in diagnostic procedures.
2 November 2021