Evolution One Series Spectrophotometers
Installing and Removing Accessories
22
Evolution One Series Spectrophotometer User Guide
Thermo Scientific
Related Topics
Installing a Reference Cell Holder
Installing a Reference Cell Holder
To install a reference cell holder
1. Open the reference compartment cover. See
Opening and Closing the Sample and
2. Grasp the cell holder by the handle and lower it into the reference compartment.
3. Insert the pin in the alignment hole and press the holder in place.
4. Close the reference compartment cover.
For descriptions of the available reference holders, see
Reference Cell Holders
or search for that
name in the Insight Pro Help system.
Note
Ensure the tubing does not obstruct the light beam or prevent the sample
compartment door from closing. If necessary, secure the tubing using the
supplied clips.
Handle